Cleaning and Maintenance
• If the cover slip sensor cannot be moved and/or you can feel glued-on material, carefully clean the
sensor using a lint-free cloth soaked in compatible solvent.
Fig. 52

7.5.11 Specimen slide outfeed

• Check the specimen slide outfeed for adhesive residue and, if necessary, clean it carefully using a lint-
free cloth soaked with compatible solvent.
7.6
Description of the required weekly cleaning measures

7.6.1 Dispenser group

We recommend rinsing the dispenser group once a week using approx. 150 ml of a solvent that is
compatible with the coverslip mountant.
Warning
Notice: Due to the high pressure when rinsing with solvent, it is necessary to wear gloves, safety goggles
and appropriate protective clothing when changing the cover slip mountant! Select an adequately sized
container to catch the rinsing residue.
• Switch off the instrument.
• Fill the additional glass bottle included in the standard scope of delivery with 150 ml of solvent.
• If a third glass bottle is needed because the replacement bottle is already filled with coverslip
mountant, you can order an additional glass bottle under order No.: 14 0464 36537.
• Unscrew the coverslip mountant bottle and remove it from the instrument.
Warning
Notice: Danger of dripping – any mountant that drips down must be picked up immediately using an
absorbent cloth.
86
1
Version 3.4, Revision M