5
Operation
Display
56
Explanation
The user has pressed the
STOP button.
The coverslipping operation
was ended. The number after
FINISHED
corresponds to the
selected
SET 1-4
program.
Action
The output magazine, if loaded, has
to be removed from the instrument.
Grabbed specimen slides or held
cover slips also must be removed
from their position by pressing
the respective RELEASE SLIDE or
RELEASE COVERSLIP button.
Pressing START reinitializes the
instrument. Afterwards, proceed
the same as when switching on
the instrument
(→ p. 42 – 5.3
Switching the instrument on or
IMPORTANT!
If the instrument is operated as a
workstation, the user must ensure
that before reinitialization, no rack
is in the transfer station.
Remove the cover slipped specimen
slides.
Version 3.4, Revision M
off).