Emergency stop using STOP
Leica CV5030
Note
Whenever the instrument is to pause only briefly
(e.g. to refill the cover slip magazine), use PAUSE
to interrupt the instrument without having to run
through a complete initialization and prime cycle.
If a just partly filled output magazine is to be
removed, press LIFT and proceed as outlined under
(→ p. 46 – 5.5 The coverslipping
In PAUSE mode, pressing and holding the
RESPOND button (>2 sec) allows you to change
coverslipping parameters or select another
parameter set (SET 1 through 4)
Button functions for
• Use the START button to continue the coverslipping
operation.
• In an emergency or if the coverslipping operation is to be
canceled quickly, press the STOP button.
• Then, the coverslipping operation is interrupted
immediately.
• The message
STOPPED
signal tone.
• Confirm the signal tone by pressing the RESPOND button.
Warning
Remove all glass parts or other objects away from
the Leica CV5030 working area after pressing
STOP. Only then may START be actuated.
• Using the RELEASE COVER SLIP and RELEASE SLIDE
buttons, you can release any cover slip that has been
picked up or specimen slide that has been gripped.
Warning
• Proceed carefully when removing a specimen slide
that has been gripped. Do not manually open the
grippers under any circumstances.
• Immediately remove all glass parts, broken
glass (e.g. cover slips) or other objects from the
instrument's working area.
Operation
operation).