Operation As St5020 - Cv5030 Workstation - Leica CV5030 Instructions For Use Manual

Robotic cover slipper
Hide thumbs Also See for CV5030:
Table of Contents

Advertisement

Note
Make sure that the connection between the staining and coverslipping machine is made using the
interface cable included in the standard scope of delivery. In addition, when programming the staining
protocols,
EXIT
will NOT be transfered into the coverslipper automatically.
• All empty specimen slide holders must be removed from the discharge chute and the drawer of the
transfer station must be closed. Then, press the START button so that the
the display. Transfer from the slide stainer to the robotic coverslipper is now possible once again.
Manual loading
Warning
Basically, we do not recommend manual loading of the Leica CV5030 in workstation operation, as this can
impair or stop the program flow of the Leica ST5010 AutoStainerXL.
• However, if it is necessary to insert a specimen slide holder into the loading bath manually for
coverslipping, the empty specimen slide holder must also be removed from the loading drawer
manually. The empty specimen slide holder is not conveyed automatically into the transfer station and
thus onto the discharge chute. Specimen slide holders from the stainer have priority for processing.
• The display shows the FINISHED message and a signal tone sounds. Confirm the signal tone with the
RESPOND button. This is followed by the
Confirm the signal tone with the RESPOND button. The display then shows the
• Do not press the START button until after the specimen slide holder is removed. This will ensure that
pending specimen slide holders from the slide stainer can be processed.
• If a specimen slide holder is waiting in the unload station of the stainer to be transferred to the
Leica CV5030, a manually loaded specimen slide holder cannot be processed.
Warning
Ensure that at this time, no specimen slide holder is manually inserted into the loading bath, as this can stop
the program flow.
Note
So that you do not disturb workstation operation, you should process specimen slide holders that
come from the slide stainer first. Only then should you process the specimen slide holders to be loaded
manually in the coverslipper.
6.2
Operation as ST5020 — CV5030 workstation
In this case, both instruments are connected to each other by a TS5025 Transfer Station.
Unlike the Leica ST5010 AutoStainerXL workstation, both instruments - the Leica ST5020 Multistainer and
Leica CV5030 - communicate with each other.
Leica CV5030
must be entered as the last step. If you do not observe this, the specimen slide holders
PAUSED
and
CHECK BATH
Workstation operation
READY
message appears on
messages and a signal tone.
PAUSED
message.
6