Key functions for instrument operation

LED
LED
LED
LED
Warning
After the output magazine reaches the output position, all specimen slides must be removed before the
coverslipping operation can be resumed.
Non-observance can lead to damage of the specimen slide.
RESPOND
Leica CV5030
START begins the coverslipping operation. This is only possible from
READY or PAUSE mode.
The READY LED is illuminated in green in READY mode and red in PAUSE
mode.
Pressing PAUSE interrupts the coverslipping operation (BUSY LED is
illuminated). Processing of the current specimen slide finishes before the
instrument switches to PAUSE mode.
The READY LED is illuminated in red afterwards.
PAUSED
appears in the display. Press START to continue the coverslipping
operation.
STOP interrupts the coverslipping operation immediately.
The ERROR is illuminated in red; the display reads
tone is output.
Press the START button to continue working. The instrument is reinitialized
when the button is pushed. Refer to
instrument on or
off), PRIME and
coverslipping
operation).
Pushing the LIFT button moves the output magazine into the removal
position.
The function is disabled while the BUSY LED is illuminated.
A signal tone can be acknowledged using the RESPOND button.