5
Operation
5.8

Button functions for programming

Press and hold for 2 s
2
Fig. 37
58
There are two empty
specimen slide holders on
the discharge chute of the
transfer station.
The capacity of the discharge
chute has been reached.
There are three empty
specimen slide holders on
the discharge chute of the
transfer station.
1
Press 1x
Remove the empty specimen slide
holders from the discharge chute of
the transfer station.
The empty specimen slide holders
absolutely must be removed
from the discharge chute of the
transfer station, otherwise no
other specimen slide holders can
be transported from the staining
machine to the Leica CV5030.
Press and hold RESPOND for approx. two
seconds to set the programs and parameter sets
the instrument uses to operate.
The individual parameter sets are set using two
submenus:
MENU A
(→ Fig. 37)
and
• Toggle between the two submenus by briefly
pressing RESPOND. The active submenu
(→ Fig. 37-1)
is displayed in the right half of the
top display line.
• There are four selectable programs
through 4) that are displayed on the upper left
half of the display (→ Fig. 37-2).
The bottom line in the display shows the
parameters in
MENU
A:
STK
- Length of the coverslip mountant stroke
VOL
- Quantity of the coverslip mountant
TYP
- Pressure in the mountant bottle
LEN
- Length correction for the coverslip mountant
stroke
MENU B
(→ Fig. 38)
(SET 1
Version 3.4, Revision M