Console Requirements; Unit Environmental Requirements; Cooling Requirements; Lighting Requirements - GE Vivid q N Service Manual

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
Section 2-2

Console Requirements

2-2-1

Unit Environmental Requirements

Table 2-2
Requirement
Operational
Storage
Transport
NOTE:
The Vivid q N portable ultrasound scanner system may be operated at an altitude of up to 3000
meters (9842 ft)
If the system has been in storage or has been transported, please see the acclimation
CAUTION
requirements before powering
page 3-2.
2-2-2

Cooling Requirements

The cooling requirement for the Vivid q N environment is 350 BTU/hr.
The cooling requirement for the SafeLock Cart environment is ~70 BTU/hr.
These figures does not include the cooling required for lights, people, or other equipment in the room.
NOTE:
Each person in the room places an additional 300 BTU/hr demand on the environmental
cooling.
2-2-3

Lighting Requirements

For system installation, updates and repairs, bright lighting is required. However, operator and patient
comfort may be optimized if the room lighting is subdued and indirect when a scan is being performed.
Therefore, a combination lighting system (dim/bright) is recommended. Keep in mind that lighting
controls and dimmers can be a source of EMI which could degrade image quality. These controls should
be selected to minimize possible interference.
2-2-4

Time and Manpower Requirements

Site preparation takes time. Begin pre-installation checks as soon as possible to allow sufficient time to
make any required changes. If possible, begin these checks as many as six weeks before system
delivery.
Only one person is required to unpack the Vivid q N ultrasound unit; at least two people must be
CAUTION
available to roll the system down the wheeling ramp. Attempts to move the system considerable
distances (or on an incline) by one person alone, could result in personal injury, and/or damage
to the system.
2-2
2
EVISION
Environmental Requirements
Temperature
o
+10 — +40
C (50 — 104
o
-20 — +60
C (-4 — 140
o
-10 — +60
C (14 — 140
Section 2-2 - Console Requirements
Relative Humidity
(non-condensing)
o
30 — 85%
F)
o
30 — 95%
F)
o
30 — 95%
F)
and/or using the system.
ON
V
N S
IVID Q
ERVICE
Air Pressure
700 — 1060 hPa
700 — 1060 hPa
700 — 1060 hPa
See "Installation Warnings" on
M
ANUAL

Hide quick links: