GE
D
FQ091019, R
IRECTION
7-6-1-3-9
Audio Test
NOTE:
For a description of this test, refer to
1.) In the Diagnostic Test window, trackball to Audio Test in the list of Interactive PC Doctor tests
available for selection from the Diagnostic Test Tree (see
name is displayed under Execute; the relevant system components to be tested are highlighted light
blue in the Schematic View (BEP, Peripherals, Keyboard), as shown below:
2.) Trackball to the Start button and press Set.
The interactive PC Dr for Windows Audio Test screen is displayed, as shown in
3.) Trackball to the Left Channel, Right Channel, Both Channels or Midi button and press Set to select
the speaker from which to play sound.
NOTE:
The Internal Speaker Beep is disabled.
7-132
2
EVISION
Figure 7-96 Audio Test
Figure 7-97 PC Doctor for Windows Audio Test
Section 7-6 - Performing Back End Diagnostics on the System
"Audio Test" on page
7-18.
Figure 7-96
V
N S
M
IVID Q
ERVICE
below) and press Set. The Test
Figure
7-97.