Table 4-7
Step
10
11
12
13
2
EVISION
External Keyboard Test (cont'd)- Recommended
Task
Perform the BUTTON ACTIVITY TEST, as follows:
a.) Change the combo box near the Front Panel button
to Reset to power up conditions and press the
Front Panel button
b.) Change the combo box near Front Panel button to
Enable transmission and press the Front Panel
button.
c.) At the Indicators frame set Indicator – ID to the All
position. Select ON and Activity.
d.) Press the Test Software button
Perform the STATUS INDICATOR TEST, as follows:
a.) Check the Battery green check box. Press Set.
b.) .Uncheck the Battery green check box. Check the
Battery yellow check box. Press Set
c.) Uncheck the Battery yellow check box. Check the
Extern DC green check box. Press Set
d.) Uncheck the Extern DC green check box. Check the
Extern DC yellow check box. Press Set
Perform the LID BUTTON TEST, as follows:
a.) Press the Diagnostic Info button.
b.) Place the test upper cover magnet close to the
Alphanumeric Keyboard Escape button.
Perform the POWER BUTTON TEST, as follows:
a.) Press the Diagnostic Info button. Press the Power
button.
b.) Press the Disconnect and Close buttons to close the
test application.
Chapter 4 - General Procedures and Functional Checks
V
IVID Q
Expected Result(s)
All 10 Activity green LEDs and KB backlights are seen
with maximal intensity
Battery green LED is On
Battery yellow LED is On
Extern DC green LED is On
Extern DC yellow LED is On
The status of the Lid switch is "Open" and changes to
"Closed "after the magnet was placed.
The status of the Power Button is "Open" and changes
to "Closed "when the power button is pressed.
N S
M
ERVICE
ANUAL
4-21