GE Vivid q N Service Manual page 518

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
8-2-4-5
Control Panel and Keyboard Installation Procedure
NOTE:
When performing the following installation procedure steps, it may be helpful to review the
appropriate movie clips previously referenced in the
Procedure.
NOTE:
When installing the Control Panel and Keyboard during other replacement procedures, Step 1
will only be required if indicated.
1) Stand the replacement Control Panel and Keyboard beside the Vivid q N system and tilt it on its
side to gain access to the cable connector underneath. Re-connect the keyboard/control panel
cable from the system to the cable connector (refer to
2.) Carefully lift the Control Panel and Keyboard into position on the Vivid q N (refer to
above). Make sure it is properly seated and take care to align the screw holes with those in the
Bottom Assembly.
3.) Close the Vivid q N LCD Display cover, making sure both front latches are securely fastened.
4.) Carefully turn the Vivid q Nover, placing it face down.
5.) Return and fasten each of the 9 screws that secure the Control Panel and Keyboard onto the
Bottom Assembly (refer to
6.) Re-install the battery as described in the
7.) Return the Handle as described in the
8.) Reconnect the AC adapter and all cables to the Vivid q N portable ultrasound scanner.
9.) Proceed to perform the following functionality tests:
-
-
-
-
-
-
-
-
-
8-12
2
EVISION
Figure
"SafeLock Cart - Grounding Continuity" on page 10-22
currently mounted on the SafeLock cart
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
"Isolated Patient Lead (Source) Leakage – Lead-to-Ground" on page 10-26
"Isolated Patient Lead (Source) Leakage – Lead-to-Lead" on page 10-27
"Isolated Patient Lead (Sink) Leakage - Isolation Test" on page 10-27
"Alphanumeric Keyboard Test" on page 4-19
"External Keyboard Test" on page 4-20
"Trackball Test" on page 4-22
"Calibration Tests" on page 7-25
Section 8-2 - External Component Replacement Procedures
Control Panel and Keyboard Removal
Figure
8-14, above).
"Battery Installation Procedure" on page
"Bearing Handle Installation Procedure" on page
V
N S
IVID Q
ERVICE
8-16, above).
Figure
8-5.
- only if the Vivid q N system is
M
ANUAL
8-15,
8-3.