Confirming Safelock Cart Voltage Configuration; System Voltage Confirmation - GE Vivid q N Service Manual

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
3-5-4

Confirming SafeLock Cart Voltage Configuration

1.) Check the safety rating plate located at the rear of the SafeLock Cart, above AC input connector.
2.) Remove the bottom cover (rear) from the SafeLock Cart to provide access to the jumper on the AC
unit. See
3.) Make sure that the jumper is set to either 100 to 120 V AC or to 220 - 240 V AC, in accordance with
the voltage rating shown on the safety rating plate (step 1).
3-5-5

System Voltage Confirmation

3-5-5-1
System Voltage Settings
Verify that the Vivid q N is set to the correct voltage.
The Vivid q N voltage settings are found on the base (underside) of the system.
For illustrations of the label with the voltage setting, see:
Table 1-5 "Main Label for Vivid q N" on page 1-19
CONNECTING THE VIVID Q N TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY
WARNING
WARNING
DESTROY THE SCANNER.
3-18
2
EVISION
Figure 3-11
below.
Figure 3-11 Jumper on AC Unit
Section 3-5 - Preparing for Installation
V
N S
M
IVID Q
ERVICE
ANUAL

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fq000040

Table of Contents