Physical Inspection - GE Vivid q N Service Manual

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
10-4-5

Physical Inspection

Table 10-9
Step
1
2
3
4
5
6
7
8
9
10
11
2
EVISION
Physical Checks
Item
Labeling
Scratches and Dents
LCD Display Cover
LCD Display
Control Panel and Keyboard
Probe Holders
SafeLock Cart
MOD
(if option installed)
DVD- CDRW Drive
Cables and Connectors
Power Cord
Chapter 10 - Care and Maintenance
Verify that all system labeling is present and in readable condition.
Inspect the system for dents, scratches or cracks.
Clean the LCD Display Cover as described in
Verify ease of up/down movement and make sure that the LCD Display cover holds
its set position when open and that there is no play on the mechanism.
Check that the cover latches open and lock correctly.
Clean the LCD Display as described in
Inspect the LCD Display for scratches and raster burns.
Verify proper operation of Contrast and Brightness controls. For details, see "LCD
Display Adjustment and Calibration" on page 3-76.
Clean the Control Panel and Keyboard as described in
Inspect the Control Panel and Keyboard. Note any damaged or missing items.
(Replace faulty components, as required).
Verify proper operation of Control Panel backlighting and TGC sliders.
Clean the probe holders as described in
Inspect the Probe Holders for cracks or damage.
Check all four SafeLock Cart wheels and castors for wear.
Check the brakes on each of the two front wheels. Make sure that pressing the
brake locks the wheel, preventing it from swivelling or rolling and therefore stopping
movement of the SafeLock Cart. Verify that releasing the brakes allows free
movement of the SafeLock Cart.
Check the SafeLock Cart raise/lower motion and verify proper operation of the
release mechanism.
Clean the drive head and media with the vendor-supplied cleaning kit. Advise the
user to repeat this often, to prevent future problems. MOD disks must be stored
away from dust and cigarette smoke. Do not use alcohol or benzene to clean the
MOD cartridge.
Clean the drive head and media with the vendor-supplied cleaning kit. Advise the
user to repeat this often, to prevent future problems. CDs must be stored away from
dust and cigarette smoke. Do not use alcohol or benzene to clean the CD drive.
Check all internal cable harnesses and connectors for wear, and secure connector
seating. Pay special attention to probe strain or bend reliefs.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation or
other deterioration, and verify continuity. Tighten the clamps that secure the power
cord to the unit and the outlet plug to the cord.
Replace the power cord and clamp, as required.