10.)Trackball to the Remote Path field and type in the IP address and path of the remote computer:
\\10.0.0.2\EVUE_IMPORT, or type in the remote computer full name \\Name\EVUE_IMPORT
(refer to
11.)Trackball to the Destination field, open the drop-down menu and select Remote Path
(\\10.0.0.2\EVUE_IMPORT), or the computer name \\Name\EVUE_IMPORT
(refer to
12.)Activate the Copy Media Player Installation check box.
NOTE:
Step 12
XP. After initial transfer of eVue, it is recommended to disable this check box.
The Vivid q N dataflow is set and ready for export/import of eVue files into a network computer.
From now on, whenever the Vivid q N is used for scanning and the Store button is pressed, another
person will be able to view the stored loop on the remote e-Vue PC (after a short transmission time).
When new loops arrive, the following message will be displayed at the bottom of the screen:
"New images have arrived - To view please refresh the screen using F5"."
3-10-8-4
Using eVue
1.) On the Vivid q N portable ultrasound scanner, in regular scanning mode, press Patient.
When the following window is displayed, select Create New Patient.
The Search/Create Patient dialog box opens.
3-134
2
EVISION
Figure
3-141).
Figure
3-141).
is recommended for the first transfer of eVue files. This step is not needed for Windows
Figure 3-142 Search/Create Patient Dialog Box
Section 3-10 - Connectivity Setup
V
N S
M
IVID Q
ERVICE
ANUAL