Radio Frequency Interface (RFI) Diagnostic Tests

7-4-3-1
Accessing the RFI Diagnostic Test Options
Access the Diagnostic Test window, as described in the Accessing the Front End Diagnostic
Options section, on page 7-25.
7-4-3-2
Full RFI Test
NOTE:
For a description of this test, refer to the information provided in
NOTE:
When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to Full RFI Test in the list of RFI tests available for selection
from the Diagnostic Test Tree (see
displayed under Execute; the relevant system component to be tested is highlighted light blue in the
Schematic View (RFI), as shown below:
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The Full RFI Test commences. While the test proceeds, the name of the test currently in progress
(Full RFI Test) is displayed in the Execute field and the current Loop # and specific test are shown
2
EVISION
Figure 7-11 Full RFI Test
Chapter 7 - Diagnostics/Troubleshooting
Figure 7-11
below) and press Set. The Full RFI Test name is
V
N S
IVID Q
ERVICE
"Full RFI Test" on page
7-11.
M
ANUAL
7-31