GE
D
FQ091019, R
IRECTION
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Fixed Voltages Test sequence is complete:
-
-
NOTE:
For this test, there is one report (Voltages) which is listed in the Reports window on the right.
By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values are within the pre-defined Max and Min limits, as
shown in
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
7-94
2
EVISION
the Diagnostic Test window displays the Finished message
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in
Figure 7-65 View Test Log - Fixed Voltages Test
Figure
7-65.
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in
If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in
Chapter 5 - Components and Function
Section 7-5 - Monitoring Tests
"DC Offset Calibration" on page
V
N S
IVID Q
Figure
7-65.
7-25, and then repeat the test.
(Theory).
M
ERVICE
ANUAL