8-2-8

LCD Display Frame Replacement Procedure

8-2-8-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-2-8-2
Time Required
10 minutes
8-2-8-3
Preparations
Shut down the Vivid q N ultrasound unit, as described in
8-2-8-4
LCD Display Frame Removal Procedure
1) With the LCD Display in the open (upright) position, use a flat screwdriver to remove the screw cap
from each of the four screws that fasten the LCD display frame in position, as shown in
below.
Figure 8-28 Removing the Screw Caps from the LCD Display Frame
2.) Loosen and remove the four screws
Figure 8-29 Removing the Securing Screws
3.) With the LCD Display in the fully open position, carefully release the LCD display frame from the
2
EVISION
Chapter 8 - Replacement Procedures
Chapter 3 - System
(Figure
8-29).
V
N S
M
IVID Q
ERVICE
ANUAL
Setup.
Figure 8-28