GE Vivid q N Service Manual page 101

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
3-7-1-4
Sony UP-D897 B/W Video Printer
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
NOTICE
devices; this may cause the system to lock-up during the boot-up procedure.
The B/W Printer connection in the different scenarios is explained below.
NOTE:
After physical connection of the printer to the Vivid q Nsystem, to configure the appropriate hot
keys to activate printing direct from the control panel, follow the instructions in
Keys to Activate Printing Direct from the Control Panel" on page
No SafeLock Cart - connection directly to the Vivid q N system
The B/W Printer can be connected to either of the USB ports (upper or lower) on the rear connectors
panel.
When connecting the B/W Printer directly to the system, it is necessary to use an additional power
source. As a safety precaution, this must be done via an isolation transformer.
The B/W Printer can be connected once the system is powered ON, or after shutdown. All software
drivers for the B/W Printer are pre-configured and installed designated to be used with the specific
B/W Printer supplied by GE.
Do not attempt to use a different type of B/W Printer (brand or model) other than the
WARNING
WARNING
B/W Printer provided by
sensitive and complex medical system. Any unauthorized peripherals may cause system failure
or damage!
No SafeLock Cart - connection via USB hub
As stated above. Safety considerations must be taken into account. The B/W Printer must be
powered via an isolation transformer. Either USB port may be used to communicate with the
device.
Vivid q N mounted on SafeLock Cart
The B/W Printer may be connected to any of the USB outlets provided on the SafeLock Cart. You
must use the additional power source to activate the B/W Printer. However, this must be powered
from the SafeLock Cart itself, via one of the isolated peripheral power outlets provided on the cart.
2
EVISION
Figure 3-21 Black and White Printer
GE Medical Systems
Chapter 3 - System Setup
V
3-22.
. The ultrasound system is an extremely
N S
M
IVID Q
ERVICE
ANUAL
"Configuring Hot
3-29