9) After reviewing the test results, proceed as follows:
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
7-6-1-3-2
Full PC Doctor Test (50 min)
NOTE:
For a description of this test, refer to the information in
7-15.
1.) In the Diagnostic Test window, trackball to Full PC Doctor Test in the list of PC Doctor tests available
for selection from the Diagnostic Test Tree (see
displayed under Execute; the relevant system components to be tested are highlighted light blue in
the Schematic View (BEP, Peripherals, Keyboard), as shown below:
2
EVISION
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in
Chapter 5 - Components and Function
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
Figure 7-82 Full PC Doctor Test
Chapter 7 - Diagnostics/Troubleshooting
(Theory).
"Full PC Doctor Test (50 min)" on page
Figure 7-82
below) and press Set. The Test name is