3-9-2

Vivid q N Configuration

3-9-2-1
Preparations
Press CONFIG (F2) and log on as Adm (see
3-9-2-2
Enter Location
a
b
Table 3-6
Step
1.
Open the Configuration Window, see
If needed, select
2.
Select the Hospital field, see (a) in
name of the hospital (max 64 characters).
3.
Select the Department field, see (b) in
name of the department (max 64 characters).
3-70
2
EVISION
Figure 3-68 Hospital and Department Name
Location Name
Task
Preparations
.
SYSTEM
Figure
3-68, and type the
Figure
3-68, and type the
Section 3-9 - Configuration
"Log On to the System as ADM" on page
.
The System Settings Window is displayed.
The 24 first characters of this name are displayed on the scanning
screen's title bar (after restart)
All 64 characters are displayed on the image properties on saved images
(after restart).
This name will be displayed on the image properties on saved images as
soon as the unit has been restarted.
V
N S
IVID Q
Expected Result(s)
M
ERVICE
ANUAL
4-10).