Peripherals - GE Vivid q N Service Manual

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
4-3-2

Peripherals

4-3-2-1
VCR Test
Table 4-12
Step
1
2
3
4
5
6
4-3-2-2
B/W and Color Printer Test
Table 4-13
Step
1
2
3
4-3-2-3
Color Printer Test
Table 4-14
Step
1
2
3
4-24
2
EVISION
VCR Test - Recommended
Task
Connect a Mitsubishi 9600 VCR and video converter to
the system, restart the system and load an SVHS
cassette.
Connect an external monitor to the VCR S-Video or
Video output.
Connect a 3S-RS probe. Reset the VCR counter and
press the VCR Record button.
Activate CW Mode and record 2 minutes of heart
scanning or a similar effect using a phantom.
Stop recording and rewind the cassette. Press the Play
button on the VCR.
B/W Printer Test - Recommended
Task
Scan a phantom or suitable interface in 2D-Mode.
Press the Freeze button.
Press the Print button.
Color Printer Test - Recommended
Task
Scan a blood vessel in 2D-Mode or a phantom
simulating a similar CFM effect, then activate CFM to
see color.
Press the Freeze button.
Press the Print P1 button.
Section 4-3 - Functional Checks
V
IVID Q
Expected Result(s)
Verify that you follow all the procedure instructions
described in
"Connecting the VCR" on page
Verify that the recording indicator (red dot) is displayed
on the VCR and the tape counter is running on the
VCR.
Record 2 minutes of CW scanning.
The recorded Image is played back.
Verify that in CW Mode both the left and right audio
channels are working, and that the played-back audio
is similar to the recorded audio in both the left and right
speakers, when applicable.
Expected Result(s)
The image scanning stops with the latest picture frozen
on the screen.
Evaluate the print quality. Adjust the brightness and/or
contrast of the B/W printer, if required. The image
displayed on the screen is printed on the B/W or color
printer, depending upon the key assignment
configuration.
Expected Result(s)
Evaluate the color print quality. Adjust the brightness
and/or contrast of the Color printer, if required.
N S
M
ERVICE
ANUAL
3-48.