GE Vivid q N Service Manual page 525

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
8.) Press Config and log on as Adm - see
9.) Select About, then click on the HwVersion tab.
10.)Verify that the correct ECG/Respiratory module Part # is displayed, as shown in
Figure 8-23 Hardware Version Verification
11.)Proceed to perform the following functionality tests:
-
-
-
-
-
2
EVISION
"Isolated Patient Lead (Source) Leakage – Lead-to-Ground" on page 10-26
"Isolated Patient Lead (Source) Leakage – Lead-to-Lead" on page 10-27
"Isolated Patient Lead (Sink) Leakage - Isolation Test" on page 10-27
"SafeLock Cart - Grounding Continuity" on page 10-22
Vivid q N system is currently mounted on the SafeLock cart
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Chapter 8 - Replacement Procedures
"Log On to the System as ADM" on page
- only if the
V
N S
M
IVID Q
ERVICE
ANUAL
4-10.
Figure
8-23.
8-19