GE Vivid q N Service Manual page 190

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
Configure Search Criteria: Constrain Query by AE Title
NOTE:
This configuration is optional. It should only be performed if the site is directing orders to the
Vivid q N that is scheduled to perform the study.
Please contact the Mitra support personnel to verify that this step is needed.
1.) In the Search Criterias dialog box, select 00400001 Scheduled Station AE Title from the Select
Tag pull-down menu.
2.) Enter the Vivid q N computer name for the Value.
3.) Click the Add to List button.
4.) Click the OK button.
Configure Search Criteria: Constrain Query by MRN
NOTE:
This configuration should be performed if the site will need to constrain the Worklist queries by
MRN (patient id)
1.) In the Search Criterias dialog box, select 00101000 Other Patient IDs from the Select Tag
pull-down menu.
2.) Enter a 9 digit number for the Value.
3.) Click the Add to List button.
4.) Click the OK button.
Assign DICOM Worklist name to Dataflow
1.) Select the Dataflow tab.
2.) From the Dataflow name pull-down menu, select the Dataflow which will be used.
3.) Remove the DICOM Worklist service currently used in the Dataflow.
4.) From the Select Service pull-down menu, select the Gateway Modality Worklist service.
5.) Click the Add button.
6.) Ensure that the service is displaying the value In for the Dir parameter and that its role is Primary.
Test Connectivity with the Vivid HL7 Gateway
The purpose of this test is to verify that the Vivid q N has connectivity with the Vivid HL7 Gateway.
1.) From either the Current or Dataflow view, expand the Dataflow incorporating the Gateway Modality
Worklist service.
2.) Expand the Gateway destination device.
3.) Select the Gateway Modality Worklist service.
4.) Click the Check button.
-
-
3-118
2
EVISION
If connectivity was achieved, then a green check should be displayed next to the name of the
Gateway's Modality Worklist service.
If connectivity was not achieved, then there will be a red X next to the name of the Gateway's
Modality Worklist service.
Section 3-10 - Connectivity Setup
V
N S
M
IVID Q
ERVICE
ANUAL

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fq000040

Table of Contents