GE Vivid q N Service Manual page 310

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
Table 5-3 RFI Board - Input DC Voltages
+3.3 V
HVP_IN
HVN_IN
+THV
-THV
+5 V_IN
-5 V_IN
+12 V_IN
1.8 V
5-3-3-7
Probe Management
5-3-3-7-1
Mechanism
The RFI Board performs the following probe management functions:
Controls the selection of a probe connector
Senses the Probe type
Senses the Scan Plane Angle on TE probes
Senses the Probe temperature
If the temperature exceeds the maximum temperature, the RFI stops scanning and disables
Tx voltage.
The temperature references are set to 41.3
If the temperature reading falls below minimum temperature, the RFI stops scanning and disables
Tx voltage.
The minimum temperature references are set to 15
indicates a probe temperature-sensing error.
5-14
2
EVISION
Input
Description
DC voltage from PS
Fixed DC high voltage +80 V
Fixed DC high voltage -80 V
Transmit high voltage
varies from 0 to +50 V
Transmit high voltage
varies from 0 to -50 V
DC voltage from PS
DC voltage from PS
DC voltage from PS
DC voltage from PS
Section 5-3 - Front End Unit
Connector - PIN#
P14 1-4
From PS
P14-9
From PS
P14-11
From PS
P14-12
From PS
P14-14
From PS
P14-16
From PS
P14-20
From PS
P14-18
From PS
P1421, 22
From PS
o
o
C and 43.0
C
o
C. A temperature reading as low as 15
V
N S
M
IVID Q
ERVICE
Connected From
ANUAL
o
C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fq000040

Table of Contents