GE
D
FQ091019, R
IRECTION
Figure 8-83 Placing the ECG Cable Protector in Position
When placing the ECG Cable Protector in position, take care not to damage the plastic.
CAUTION
IMPORTANT: Perform the next steps in the correct order. When securing the ECG Cable Protector, first fasten the rear
securing screw, then the front securing screw.
3.) Return and fasten the rear securing screw
Figure 8-84 Fastening the Rear Securing Screw
Figure 8-85
2
EVISION
illustrates the ECG Cable Protector secured in position on the SafeLock Cart.
Chapter 8 - Replacement Procedures
(Figure
8-84), then the front securing screw.
V
N S
M
IVID Q
ERVICE
ANUAL
8-71