Data Sheet for ECG Current Leakage
The test passes when all readings measure less than the values shown in the tables below. Record all
data.
Table 10-17 Maximum Allowance Limit for ECG Lead Leakage Test to Ground
Patient Lead-to-Ground Current Leakage Test
Patient Lead-to-Lead Current Leakage Test
Table 10-18 Maximum Allowance Limit for ECG Lead Isolation Sink Test
Table 10-19 Typical Data Sheet for ECG Lead-to-Lead Test
ECG
Power
ON
ON
ON
ON
OFF
OFF
OFF
OFF
10-28
2
EVISION
and
Patient Lead Isolation Current Test
Tester
Tester
Polarity
Ground
Switch
Switch
NORM
CLOSED
REVERSE
CLOSED
NORM
OPEN
REVERSE
OPEN
NORM
CLOSED
REVERSE
CLOSED
NORM
OPEN
REVERSE
OPEN
Section 10-6 - Electrical Safety Tests
AC Power
GROUND
Source
115V
220/240V
AC Power
Source
115V
220/240V
Tester Lead Selector
RL
RA
V
N S
IVID Q
ERVICE
Maximum
Allowance Limit
GROUND
OPEN
CLOSED
10uA
10uA
500uA
10uA
Maximum
Allowance Limit
20uA
5mA
LA
LL
M
ANUAL
C