FQ091019, R
IRECTION
3.) Lift off and remove the defective knob - see
Figure 8-20 ATGC Rotary Knob Replacement
8-2-4-5-8
ATGC Rotary Knob Installation Procedure
1) Carefully place a new ATGC Rotary Knob in position on the keyboard, taking care to place the
plastic alignment pin in the correct location.
2.) Gently push the ATGC Rotary Knob down until it snaps into position.
8-16
2
EVISION
Section 8-2 - External Component Replacement Procedures
Figure 8-20
below.
V
N S
M
IVID Q
ERVICE
ANUAL