GE Vivid q N Service Manual page 604

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
8-4-19-4
Rear Wheel Installation Procedure
1) Place the new rear wheel on the SafeLock Cart chassis in the same position as that of the
previously-removed wheel.
2) Return and fasten the four bolts (previously removed - see
is secured firmly to the chassis.
3) Stand the SafeLock Cart upright.
4.) Return the Vivid q N and all peripherals to their appropriate positions on the SafeLock Cart.
5.) Re-connect the AC adapter and all cables and peripherals to the Vivid q N portable ultrasound
scanner.
6.) Re-connect mains power to the AC adapter.
7) Proceed to perform the following functionality tests:
-
-
8-98
2
EVISION
"SafeLock Cart - Grounding Continuity" on page 10-22
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Section 8-4 - SafeLock Cart Components Replacement
V
N S
IVID Q
Figure
8-111) and make sure that wheel
M
ERVICE
ANUAL