GE Vivid q N Service Manual page 440

Hide thumbs Also See for Vivid q N:
Table of Contents

Advertisement

GE
D
FQ091019, R
IRECTION
7) When finished reviewing the first report, trackball to the Exit button and press Set to close the View
Report window.
8) Select the second report (HPF600); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in
9) When done, trackball to the Exit button and press Set to close the View Report window.
10) After reviewing the test results, proceed as follows:
11.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
12.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
7-82
2
EVISION
Figure 7-54 View Test Log - ACW HPF Test - 600
Re-boot the system (this is obligatory).
Perform calibration, as described in
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in
Chapter 5 - Components and Function
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance
Section 7-4 - Performing Front End (FE) Diagnostics
"DC Offset Calibration" on page
(Theory).
V
N S
M
IVID Q
ERVICE
ANUAL
Figure
7-54.
7-25.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fq000040

Table of Contents