System Calibration Procedure (Autocal)
PD1000-60002 Protection Probe Calibration – I
Hardware Setup: Refer to Figure 12 below.
Figure 12 PD1000-60002 Protection Probe Calibration Setup
Procedure:
After setting up the hardware as described above, click Start to perform the
calibration and then continue to the next step.
Keysight Double-Pulse Test System Installation Guide
– Connect the PD1000-60002 Protection Probe to the Oscilloscope Channel
2.
– Connect the BNC to Banana Adapter to the BNC to SMA cable.
– Connect the SMA end of the cable to the SMA connector on the
PD1000-60002 Protection Probe. Use the 0.56 N-m (5 lb-in) 5/16 in. SMA
Break-over torque wrench (provided with the PD1500A system) to tighten
the cable to the PD1000-60002 Protection Probe.
– Connect the banana plug cable to Channel 1 Force terminals of the
B2902A SMU (2-wire measurement).
– The Safety Enclosure hood may be left open for this calibration step.
Use the 0.56 N-m (5 lb-in) 5/16 in. SMA Break-over torque wrench
(provided with the PD1500A system) to tighten the cable to the
PD1000-60002 Protection Probe. Failure to do so might cause
extensive ringing in the DPT measurements. See appendix for SMA
connector care and connection information.
PD1500A Double-Pulse Test System Calibration
/I
D
C
1251-2277
Banana - BNC Adapter
See Note below for
SMA connection
information.
PD1000-60002
Protection Probe
8121-2816
BNC - SMA
Cable
71