PD1500A Double-Pulse Test System Calibration
3 Connect the oscilloscope probes to the Deskew section of the Test Fixture.
The N2873A V
Probe
CLAMP
would connect here for
Deskew instead of the
N2819A Differential Probe
Figure 14 Oscilloscope Probes Connected to Deskew Fixture
Procedure:
1 After setting up the oscilloscope probes in the Deskew fixture, click Start to
2 When complete, return the oscilloscope probes to the appropriate connectors
3 Return the SMA cable from the Cal (Deskew) Input SMA connector to the Low
74
IMPORTANT: The N2918A Differential Probe must face to the left; the
Keysight logo must face toward the V
perform Deskew calibration.
on the DUT and Test Modules.
Side SMA connector on the Test Fixture.
IMPORTANT: After calibration and before running any Double-Pulse
Test, make certain the Low-Side Gate Drive Modules have the V/NC
jumper installed and the I/NO jumper removed. Remove all cables
from the B2902A SMU.
Oscilloscope Probe Deskew
N2819A I
Differential Probe
G
N2873A V
GS
10076A High Voltage V
I
/I
(PD1000-60002) Probe Cable
D
C
The BNC connector (ID/IC) on
the far right is for the
PD1000-60002 Oscilloscope
Protection probe cable that
would normally connect to the
Coaxial Shunt Resistor on the
DUT Test Module.
probe connector.
CLAMP
Keysight Double-Pulse Test System Installation Guide
/V
Probe
GE
/V
Probe
DS
CE