Step 4: Obtain and Install Coaxial Shunt Resistor De-Embedding File

Install the De-Embedding Transfer File on the DSOS104 Oscilloscope

1 From the temporary location where you unzipped the de-embedding file, copy
Note: On the DSOS104 Oscilloscope, the file path will show as:
C:\Users\Public\Public Documents\Infiniium\Filters. However, this is an alias
path. Use the path: C:\Users\Public\Documents\Infiniium\Filters in the Shunt
Settings De-embedding File path in the Control Software.
2 In the PD1000A Control Software, type the complete path and .tf2 file name
Keysight Double-Pulse Test System Installation Guide
For new PD1500A systems, the de-embedding file was installed in
the DSOS104 oscilloscope at the factory with the file name:
MY5840xxxx.tf2
where xxx is the serial number labeled on the shunt itself.
the .tf2 file to the DSOS104 Oscilloscope. Place it in the folder: C:\Users\
Public\Documents\Infiniium\Filters.
into the Double-Pulse Measurement Control > Settings > Instrument Configuration >
De-embedding File field. Refer to the PD1500A Control Software Guide for
additional information.
The oscilloscope channel you specify for the Output Current I
the DPT Control Software Hardware Configuration screen is the
channel that the Coaxial Shunt De-embedding is applied to. Refer
to the PD1500A Control Software Guide for detailed information.