Coaxial Shunt Resistor Characterization; Install The De-Embedding Transfer File On The Dsos104 Oscilloscope - Keysight PD1500A System Installation Manual

Double-pulse test rack and safety enclosure
Hide thumbs Also See for PD1500A:
Table of Contents

Advertisement

PD1500A Double-Pulse Test System Calibration

Coaxial Shunt Resistor Characterization

There is no separate calibration for the PD1000-61901 Coaxial Shunt Resistor.
The De-embedding characterization file installed in the DSOS104 Oscilloscope
provides the necessary accuracy.
After the shunt is characterized, Keysight will return the shunt to you. This
provides the latest characterization .tf2 file. Download the file from the Keysight
Infoline website and install it in the DSOS104 oscilloscope.

Install the De-Embedding Transfer File on the DSOS104 Oscilloscope

1 From the temporary location where you unzipped the de-embedding file, copy
2 In the PD1000A Control Software, type the .tf2 file name into the Double-Pulse
where xxxx is the serial number labeled on the shunt itself.
72
The PD1000-61901 Coaxial Shunt Resistor should be removed
from the DUT Test Module and shipped to a Keysight Service
Center for characterization. This should be done annually.
the .tf2 file to the DSOS104 Oscilloscope. Place it in the folder:
C:\Users\ Public\Documents\Infiniium\Filters.
On the DSOS104 Oscilloscope, the file path will show as:
C:\Users\Public\Public Documents\Infiniium\Filters. However, this is
an alias path. Use the path: C:\Users\Public\Documents\Infiniium\
Filters in the Shunt Settings De-embedding File path in the Control
Software.
Measurement Control > Settings > Instrument Configuration > De-embedding File
field. For new PD1500A systems, the de-embedding file was installed in the
DSOS104 oscilloscope at the factory with the file name:
MY5840_xxxx.tf2
The oscilloscope channel you specify for the Output Current ID/IC in
the DPT Control Software Hardware Configuration screen is the
channel that the Coaxial Shunt De-embedding is applied to.
System Calibration Procedure (Autocal)