Turn-on and Software Installation

Step 4: Obtain and Install Coaxial Shunt Resistor De-Embedding File

For maximum DPT accuracy, the Coaxial Shunt Resistor's De-embedding File
(also known as a Transfer File) must be installed in the DSOS104 Oscilloscope.
Recharacterization of the Coaxial Shunt should be performed annually and the
updated De-embedding file installed in the oscilloscope. The Coaxial Shunt
Resistor's De-Embedding file has to do with maximizing the Shunt Resistor's
bandwidth and requires configuration on the DSOS104A Oscilloscope.

Verify D9020ASIA License on the DSOS104 Oscilloscope

Keysight's D9020ASIA Advanced Signal Integrity Software license should be
installed on the DSOS104 Oscilloscope and the license activated at the factory.
Verify the installation and activation with the following procedure. License
activation instructions and certificate are provided with your DSOS104
Oscilloscope.
To verify that the license is installed and activated:
1 On the oscilloscope select: Help > About Infiniium.
2 Look for InfiniiSim Advanced.
Figure 4
54

Step 4: Obtain and Install Coaxial Shunt Resistor De-Embedding File

PD1000A De-Embedding License is Activated
Keysight Double-Pulse Test System Installation Guide