System Calibration Procedure (Autocal)
Gate Current Calibration – I
Hardware Setup:
Low-Side Gate
Drive Module
Figure 8
Setup for Gate Current Calibration
Procedure:
After setting up the hardware as described above, click Start to perform the
calibration and continue to the Gate Resistance Calibration.
Keysight Double-Pulse Test System Installation Guide
G
Refer to Figure 8 below.
– Install the I/NO Jumper on the Low-Side Gate Drive Module.
– Install the V/NC Jumper on the Low-Side Gate Drive Module.
– Connect the N2819A Differential Oscilloscope Probe to the I
on the Low-Side Gate Drive Module. Carefully observe polarity when
installing probe.
– Connect the U1168B Calibration Cable between the Channel 1 Force
terminals of the B2902A SMU (2-wire measurement) and the + and - CAL
eyelet terminals on the Low-Side Gate Drive Module.
– The Safety Enclosure hood may be left open.
N2819A
Differential
Probe
A strip of 10 header jumpers (Keysight part number 1258-0225)
were supplied with your PD1500A for use during calibration.
PD1500A Double-Pulse Test System Calibration
Probe
U1168B
Standard
Probe Kit
mini grabbers
pins on the
G
I/NO (Normally Open) Jumper.
V/NC (Normally Closed)
Jumper
67