Oscilloscope Probe Deskew

Oscilloscope Probe Deskew

Connect Probes to Deskew Fixture

After the Oscilloscope Probe Compensation and System Calibration, the last step
is to run Deskew. Deskew has the greatest impact on both On and Off Delay
times and hence affects the total Switching Time. Both On and Off Switching
Energies are also affected.
Hardware Setup
Because the oscilloscope has only four input channels and the DPT test fixture
uses only four oscilloscope probes at a time, then only four probes may be
deskewed at one time. Typically, either the N2819A Differential Probe (for
measuring I
they both used at the same time. Deskew the probes that you will be using in
your test. Figure 14 below shows the typical configuration for Deskewing the
oscilloscope probes.
1 Locate the Deskew oscilloscope probe connectors on the main Test Fixture.
2 Remove the SMA cable from the Low Side SMA connector on the Test Fixture
Connect Low-Side (Channel 1) of Waveform Generator to Cal (Deskew) Input
Figure 13 33512B Function Generator Connected to Deskew Fixture
Keysight Double-Pulse Test System Installation Guide
) or the N2873A (V
G
See Figure 13 below.
and connect it to the Cal (Deskew) Input SMA connector.
PD1500A Double-Pulse Test System Calibration
) probe are used in a DPT test; rarely are
CLAMP
73