System Calibration Procedure (Autocal)
High Voltage
Hardware Setup:
PD1000-66843
High-Side Gate
Drive Module (0 W
Gate Resistor)
10076C High
Voltage Probe
I
nstall 1200 V
FET in
High-Side
position
Figure 10 Setup for High Voltage Probe Calibration
Procedure:
After setting up the hardware as described above, click Start to perform the
calibration.
Keysight Double-Pulse Test System Installation Guide
Probe Calibration
– Install a PD1000-66843 High-Side Gate Drive Module (0 Gate Resistor)
– Install a 1200 V FET (for example, the SCT2080KE FET supplied with the
PD1500A) in the High-Side position of the DUT module.
– Remove any FET in the Low-Side position of the DUT Module.
– Remove (do not install) the Low-Side Gate Drive Module.
– Connect the 10076C High Voltage Probe to the DUT Module.
– Remove all other cables and oscilloscope probes.
Safety Enclosure hood must be closed for this test.
PD1500A Double-Pulse Test System Calibration
– V
/V
Probe
DS
CE