Connect Heater And Thermocouple To Device Under Test (Optional); Second Thermocouple Jack - Keysight PD1500A System Installation Manual

Double-pulse test rack and safety enclosure
Hide thumbs Also See for PD1500A:
Table of Contents

Advertisement

Connections on Front of Test Fixture

Connect Heater and Thermocouple to Device Under Test (Optional)

Keysight provides a device heater, integrated thermocouple and clamps
(PD1000-62001) to allow you to heat your device (ambient to +175 °C
DPT system measures its characteristics. The device heater comes with an
integrated Type K Thermocouple. Use the Device Clamps to attach the heater
and Thermocouple to the device being tested.
Device Heater
with integrated
Thermocouple
Figure 10 Device Heater with Thermocouple and Device Clamps
Refer to Chapters 2 (TO-247 devices) and 3 (SMD devices) for instructions to
attach the heater and thermocouple to the DUT.

Second Thermocouple Jack

A second Thermocouple jack is present on the left side of the DPT test fixture.
This jack is not functional and is reserved for possible future use only.
1 Requires Software version 2020.820 or later. Ambient to +175
Keysight Double-Pulse Test System Installation Guide
Use of the device heater and thermocouple is optional in the DPT
test system. If you want to use the heater and thermocouple, they
must be connected to the Test Fixture before powering-up the
PD1500A system. Reboot the PD15000A if the heater is not
detected.
°
+150
C for SMD devices.
Install Test Fixture in the Safety Enclosure Hood
Clamp for TO-247
and SMD devices
°
C for TO-247 device packages only; ambient to
1
) as the
Power Plug for
Device Heater
Thermocouple
Plug
43