PD1500A Double-Pulse Test System Calibration
Clamp Circuit Probe
Hardware Setup:
N2873A V
GS
Clamp Voltage
Probe
PD1000-66805
Clamp Module
Figure 11 Setup for Clamp Circuit Probe Calibration
Procedure:
After setting up the hardware as described above, click Start to perform the
calibration and then continue to the next step.
70
Calibration
See Figure 11 below.
– Install the PD1000-66805 Clamp Circuit Module in the PD1500A Test
Fixture. If you are not using or calibrating the Clamp Module, press the Skip
button to skip this step and proceed to the Protection Probe Calibration.
– Connect the N2873A V
Clamp Circuit Module.
– Using the U1168B Standard Probe Kit, connect banana plugs to Channel 1,
Force output of the B2902A SMU.
– On the other end of the test cables, attach two mini clips (one red, one
black).
– Connect the mini clips to the CAL pins on the Clamp Circuit Module.
– The Safety Enclosure hood may be left open.
/V
GE
System Calibration Procedure (Autocal)
– V
Probe
CLAMP
/V
Clamp Voltage Probe to the top of the
GS
GE
Keysight Double-Pulse Test System Installation Guide
U1168B Standard
Probe Kit with mini
clips
B2902A High Side
Force terminal
connects to the + pin
on the Clamp Circuit
Module. B2902A Low
Side Force terminal
connects to the – pin
on the Clamp Circuit
Module.