Double-Pulse Test Rack and Safety Enclosure Location

Double-Pulse Test Rack and Safety Enclosure Location
The DPT Rack and Safety Enclosure is a free-standing rack with test instruments
installed and a Safety Hood on top. It is the customer's responsibility to consider
the anticipated location and requirements to ensure safety and optimize access
to test and control instrumentation.
The following figure shows a typical layout for the Rack and Safety Enclosure.
You should allow a minimum one meter of space in front of the rack for operator
access, one meter of space behind the system for service and 61 cm on either
side of the rack for sufficient air flow for cooling and ventilation.
Allow a minimum of one meter above the Rack and Safety Enclosure for opening
the Safety Enclosure hood.
Keysight Double-Pulse Test System Installation Guide
Instrument and
Service Access Area
(minimum required for
ventilation)
100 cm (39.4 in)
Keysight
Double Pulse Rack
and Safety Enclosure
(top view)
50 cm
(20 in)
Rack Front
Operator Area
(for example purposes only.
The operator does not have to
be in front of the test system.)
The Air Intake fans on the rack sides are temperature controlled.
They turn on only when the temperature in the lower portion of
the rack exceeds approximately 30 ° C.
Site Preparation
50 cm
(20 in)
17