Install Test Fixture in the Safety Enclosure Hood

Placing the Test Fixture in the Safety Enclosure Hood

1 From the front of the rack, open the Safety Hood.
2 Carefully place the Test Fixture on the Work Surface, just to the left of the
3 Ensure the fixture is positioned so that you can close the safety hood.
Figure 6
Placing the Test Fixture inside the Safety Hood
40
oscilloscope.
There is a fan on the right side of the Test Fixture. Make certain
there is adequate air flow. At least 2 inches between the Test
Fixture and the DSOS104 Oscilloscope.

Placing the Test Fixture in the Safety Enclosure Hood

Keysight Double-Pulse Test System Installation Guide