PD1500A Double-Pulse Test System Calibration
4 Adjust so that the square wave on the oscilloscope screen looks square.

Offset Compensation for the N2819A Differential Probe

The N2819A Active Differential Probe also requires offset compensation. This is
described below and in the
trimmer tool supplied with the probe to perform the offset zero calibration.
Probe Adjustment
Tool
1 Connect the probe to an oscilloscope channel input. Turn on the oscilloscope
2 Using the probe's 5 cm Extension Leads and the Pincer or Micro IC Clip
3 Press Control > Autoscale on the oscilloscope.
62
Under Compensated
Due to the 100:1 divider of the 10076C High Voltage Probe, the
signal will appear to be very noisy. Therefore, it is beneficial to
enable averaging in the Setup > Acquisition menu for suppressing
this noise. Setting the number of averages to 16 will result in a
clean signal.
N2819A Offset Adjustment
and wait for a minimum of 20 minutes to allow the oscilloscope and probe to
warm up.
connection accessories, short the + and - inputs together.
Pincer or Micro IC Clips
Oscilloscope Probe Offset Compensation
Over Compensated
N2818/9A Differential Probe User
5 cm Extension Leads
Keysight Double-Pulse Test System Installation Guide
Optimum
Guide. Use the