DC Attenuation/Offset Calibration Using N7023A Power Rail Probe
Browser
1
Turn on the oscilloscope and connect the N7023A power rail probe browser with
the N7020A probe attached to one of the oscilloscope's input channels.
2
Allow the oscilloscope and probe to warm up for 20 minutes.
3
If the oscilloscope needs calibration, perform a user calibration before the probe
calibration. On the oscilloscope, click Utilities > Calibration.
4
Connect the ground lugs and the output of the N7023A power rail probe browser
to the oscilloscope's front-panel PROBE COMP connector.
5
On the oscilloscope, click Setup > Probes.
6
In the Probe Calibration dialog box, select the tab representing the channel that
has the probe attached.
7
In the dialog box, select the type of calibration. Click Start and follow the
instructions shown on the oscilloscope.
N7020A User's Guide
3