Using InfiniiMax III+ Series Probes
To Probe Single-Ended Signals with a Differential Head
Using a differential probe head rather than a single-ended probe head to probe
single-ended signals results in the advantages of higher bandwidth and increased
accuracy. Also no offset range is sacrificed when using a differential probe head,
as any supplied offset is applied only to the probe head's plus side. To learn more
about applying a DC offset when probing, refer to
Large DC Level"
If possible, orient the probe head vertically as shown in
probe head flat over a single-ended signal will cause coupling to the tip that can
degrade the performance. Notice that the head's "–" lead is connected to ground.
Figure 18 Connecting a Differential Head to a Single-Ended Signal
26
on page 27.
"To Measure Small Signals on a
Figure
18. Laying the
InfiniiMax III+ Series Probes User's Guide