B. Braun Dialog+ Service Manual page 449

Sw 9.xx
Table of Contents

Advertisement

+
Dialog
8 . A p p e n d i x
SW 9.xx
+
The Dialog
machine is intended for use in the electromagnetic environment specified below. The customer or the user of the Dialog
environment.
Immunity test
Immunity test
Immunity test
Immunity test
Electrostatic discharge (ESD) acc.
to IEC 61000-4-2
Electrical fast transient/burst acc.
to IEC 61000-4-4
Surge acc. to IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines acc. to IEC
61000-4-11
Power frequency (50/60 Hz)
magnetic field acc. to IEC 61000-
4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
BA-TE-DE08C M.KAY -
Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd
Electromagnetic Immunity, Table 202
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601 test level
IEC 60601 test level
IEC 60601 test level
IEC 60601 test level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for power supply
± 2 kV for power supply
lines
lines
± 1 kV for input/output for
± 1 kV for input/output for
lines
lines
± 1 kV line(s) to line(s)
± 1 kV differential mode
± 2 kV line(s) to earth
± 2 kV common mode
< 5 % U
< 5 % U
T
(> 95 % dip in U
)
(> 95 % dip in U
T
for 0.5 cycle)
for 0.5 cycle)
40 % U
40 % U
T
(60 % dip in U
)
(60 % dip in U
T
for 5 cycles
for 5 cycles
70 % U
70% U
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
< 5 % U
< 5 % U
T
(>95 % dip in U
)
(>95 % dip in U
T
for 5 s
for 5 s
3 A/m
3 A/m
Compliance level
Compliance level
Compliance level
Compliance level
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of
T
+
)
the Dialog
machine requires continued operation during power mains interruptions, it is
T
recommended that the Dialog
battery.
T
)
T
T
)
T
T
)
T
The power frequency magnetic field should be measured in the intended installation location to
assure that it is sufficiently low.
1 / 2 0 1 0
+
machine should assure that it is used in such an
Electromagnetic environment – guidance
Electromagnetic environment – guidance
Electromagnetic environment – guidance
Electromagnetic environment – guidance
+
machine be powered from an uninterruptible power supply or a
8-5
B. Braun Avitum AG

Hide quick links:

Advertisement

Chapters

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Dialog+ and is the answer not in the manual?

Table of Contents

Save PDF