4 . T S M S e r v i c e P r o g r a m
• Manometer or pressure measurement instrument
• Syringe
• Y piece
• Tubing and tubing clamp
1 / 2 0 1 0
1. Activate the LL Manual Calibration menu with
the LL Manual Calibration key in the Manual
Test and Calibration menu.
2. Activate the Pressure Sensors Water Side menu
by pressing on the line Calibration 2.3 Pressure
Sensors Water Side. The following menu is
displayed.
4 - 126