Do you have a question about the Dialog+ and is the answer not in the manual?
Questions and answers
Eduardo Felipe de Paula
January 22, 2025
ola tenho varias placas da bbraum que gera a imagem pro monitor modelo: BBM-LX800-A1R ele simplesmente ao ser ligada não bipa, e consequetemente nao funciona nada! so ascende um led na placa e mais nada. gostaria de saber se o defeito pode ser bios, memoria chipset, ou o processador que no caso e o AMD GEODE. desde de ja agradeço
1 comments:
Mr. Anderson
February 10, 2025
The cause of the B. Braun Dialog+ board not beeping or functioning when connected to a BBM-LX800-A1R monitor, with only one LED lighting up, could potentially be related to the BIOS, memory, chipset, or AMD GEODE processor. However, based on the context provided, the most likely causes to investigate include:
1. BIOS Issue: Ensure the BIOS is functioning properly and has the correct version. Refer to the procedure for checking the BIOS version as outlined in the context (e.g., pressing the TAB key during startup to display the BIOS version). A corrupted or incompatible BIOS could prevent the board from functioning.
2. Memory Issue: Faulty or improperly seated memory modules can cause similar symptoms. Verify that the memory is correctly installed and functional.
3. Chipset Issue: A malfunction or damage in the motherboard chipset could also result in the board failing to operate.
4. AMD GEODE Processor Issue: If the processor is damaged or not seated correctly, it could prevent the board from functioning.
Careful diagnostics are needed to isolate the specific cause.
Page 1
Dialog SW 9.xx S e r v i c e M a n u a l E n g l i s h Edition 1 -2010 M.KAY Dialog+ SW9xx SM EN 1-2010...
Page 2
1 / 2 0 1 0 SW 9.xx Contact your Local B. Braun Representative for Service Support 34209 Melsungen B. Braun Avitum AG Germany Tel. No.: +49 5661 713500 (Günter Nissen) E-Mail: guenter.nissen@bbraun.com B. Braun Avitum AG BA-TE-DE08C M.KAY -...
Page 3
S e rvi ce Manu al s w i th a re gi s tr a ti on num be r a re inc lud ed in th e u p da te s erv ic e! B. Braun Avitum AG BA-TE-DE08C M.KAY -...
Page 4
1 / 2 0 1 0 SW 9.xx Copyright This document is the property of B. Braun Avitum AG with all rights reserved. Commissioning and Service Only trained personnel must service the Dialog+, i.e. repair, maintenance, software installation, firmware update, retrofitting and commissioning of the Dialog+.
Page 5
Flow, Wiring and Tubing Diagrams Spare Parts List Appendix ESD/EMC Information 8.1.1 Electrostatic Discharge ESD 8.1.2 Electromagnetic Compatibility EMC Technical Information Assembly Instructions Field Service Information Instruction Leaflets Edition/Updates Service Manual B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_TOC_1-2010.doc/pdf <100329> yymmdd...
Page 6
Calibration The symbol appears for necessary calibration measures. Calibration Equipment/Tools symbol gives remarks necessary calibration equipment/tools, i.e. during a calibration, disassembly or assembly. B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_TOC_1-2010.doc/pdf <100329> yymmdd...
Page 7
Perform regular technical safety inspections as described in chapter 5 of this Technical Safety Inspection service manual to ensure the safety of the machine. An instructions for use can be ordered at your local B. Braun representative or Instructions for Use dealer.
Page 8
1 - 1 SW 9.xx Copyright This document is the property of B. Braun Avitum AG with all rights reserved. Commissioning and Service Only trained personnel must service the Dialog+, i.e. repair, maintenance, software installation, firmware update, retrofitting and commissioning of the Dialog+.
Page 9
(fasten with a single ear clamp 19.5 and a tubing clamp 12- 20 mm) • Tubing length: approx. 3 m • Drain height: max. 80 cm 8. Emergency Power Supply/Battery Option Fig.: Dialog+ Rear View B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd...
Page 10
1 / 2 0 1 0 1 - 3 SW 9.xx Table of Contents Page Commissioning Check List 1.1.1 Measurement Circuits for the Measurement of the Electrical Safety According to EN 62353/60601-1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd...
Page 11
..................................Operating Hours: ........h Inventory No.: ..............SW Version: ........... Manufacturer: B. Braun Avitum AG 34209 Melsungen, Germany Check List Note: Text in { } brackets is information for the execution of the check list! Visual Inspection {Machine: clean/complete; no damages/moisture influences or loose assemblies; no moveable parts touching tubings or wires; casters are moveable;...
Page 12
Next Inspection Date: ................................Name Service Technician: Name of Company: The commissioning was performed and the machine was hand over to the responsible ................... organisation (user)..............................Date/Signature B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd...
Page 13
The test current must be measured in both directions. Fig. 1: Protective Earth Resistance Equipment Leakage Current: • Differential Measurement L(N) N(L) Fig. 2: Equipment Leakage Current Patient Leakage Current Fig. 3: Patient Leakage Current B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd...
All sensor data are sent separately, via two serial bus systems, to the supervisor and controller via the analog board to the digital board. The actuators, motors and valves are driven via the power board valves PBV and power board motors PBM. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
Page 17
2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 - 3 SW 9.xx For Software ≥ ≥ ≥ ≥ 9.xx For Software 9.xx For Software For Software 9.xx 9.xx B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
NSVB, BICSS, KSS, MSBK1/2, RDV, SAD, BPADS, BPVDS, Service Switch: BL (Controller/Supervisor Sensor Analog Board) Position 0: Therapy Mode Position 2: TSM Service Program Mode Position 3: Software Installation/Update Mode B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
• Monitoring of the blood side pressures (pressure sensors PA, PV, PBE and PBS) • Test of the blood side pressure sensors in preparation (3/2 way solenoid valve VBT) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 - 24 SW 9.xx 2.19 Flow Diagrams 2.19.1 Dialog+ B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 - 25 SW 9.xx 2.19.2 Dialog+ with BIC Option and DF Filter Option B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
Dialog 2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 - 26 SW 9.xx 2.19.3 Dialog+ HDF Online B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
The cold inlet water is warmed up via the heat exchanger WT. Thereby the heat consumption to heat up the water is reduced. VVB* VBKS RVBO* Fig. : Water Inlet with Water Block B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
FPE. If the set pressure is reached RVFPE is opened and the fluid can circulate. RVFPE TSD-S TSDE TSBIC BICLF ENDLF ENDLF-S VDEBK1 VEBK1 BICP MSBK1 VDEBK2 VEBK2 MSBK2 Fig. : Dialysate Processing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
VLA is opened and air is removed with positive pressure from LA. VLA is closed once air is not detected anymore by LAFS. The balance chamber is switched to normal operation. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
Page 50
C and conductivity is below 0.1 mS/cm. Then the machine gives a message that asks the user to confirm that the machine is free of disinfectant and the disinfection is completed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 - 43 SW 9.xx 2.22.2 System Integration SMPS-MC Fig. : System Integration B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 - 44 SW 9.xx 2.22.3 Component Layout SMPS-MC Fig. : Layout SMPS-MC B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
2 x 12 V, 7 Ah 10 AT (5x20), breaking capacity 1500 A, 150 Vdc Fuse for Battery Jumper X104 Charge Current for Batteries Default: 50 % (right position) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
±12 VD: Digital +24 VGB: Switched blood side +24 VGD: Switched dialysate side +24 VL: Power (uncontrolled) GNDAN: GND analog GNDD: GND digital GNDH: GND aux. voltage GNDL: GND power B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
Page 60
A DC/DC converter generates the –12 V voltage from the 24 V. The soldered fuse F500 disconnects the 24 V in case of a short circuit in the DC/DC converter. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
+24 VGD 3.15 A (TR5) P4/15 P12/14 Note: Soldered fuses (e.g. F300 or F500) must not be changed if they are defective, i.e. the complete SMPS-MC must be exchanged. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
= 10 s after the switch-on of the switch mode power supply. The watchdog has no safety function and is therefore not tested before the therapy starts. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd...
Page 63
3-25 3.3.8.11 Pressure Reduction VABK with UFP 3-26 3.3.8.12 VEBK Test 3-27 3.3.8.13 Pressure Reduction 3-28 3.3.9 Integrity Test HDF Online 3-29 3.3.10 UF Pump 3-31 3.3.11 Conductivity 3-32 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 64
3-55 3.5.9 Test Run 3-55 3-56 Field Service Utilities FSU 3.6.1 FSU Directories and Subdirectories 3-56 3.6.2 Start Field Service Utility FSU 3-57 3.6.3 Set-Up FSU Main Menu 3-58 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 65
Measures after Repair 3.8.1 Repair Matrix 3-103 3.8.2 Tests and Measures to be Executed 3-104 3.8.2.1 Install Software 3-104 3.8.2.2 Check System Configuration, Production Report and Set if 3-104 Necessary B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 66
Test Run UF Comparison Measurement 3-105 3.8.2.6 Test Run 3-105 3.8.2.7 Electrical Safety Check 3-106 3.8.2.8 Perform Disinfection after Repair 3-106 3.8.2.9 Document the Executed Activities 3-106 3.8.3 Execution Protocol 3-107 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
If an error is detected during the MPU test the following alarms are displayed: In Therapy – Therapy Selection: • Hardware error RAM/ROM (SUP) In TSM Service Program: • Error in MPU test B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Format of LLS Test Codes: (Value ≥ 10000) LLS test code to LLC S S S Performed test step Example: 11104 Blood leak detector test: measure level for red and green B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Pressure sensors blood side: Equality test for blood side pressure sensors Pressure sensors blood side: Blood side pressure test Disinfection Valve VD Pressure test substitution line S-Online HDF online B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Set set point for heater to self test code #112 2720x Wait for feedback of test result via BIOS (timeout 15 s) 2730x Set set point of heater to 0 2799x Completed VGD ON VGB ON B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Activate buzzer after approx. 1 s for 250 ms 28003 2820x Wait for feedback of test result via BIOS (timeout 15 s) 2830x Set set point of heater to 0 2899x Completed VGD ON VGB ON B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Acknowledge with switch-off VGD 16022 – 16027 1640x Wait for D24OFF (timeout 60 s since start of code 200) VGD ON 16028 1650x Wait for D24ON (timeout 10 s) 1699 Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
∆PDA < 25 mmHg, 12077 VBP ---------------------------------- opened VDA, VDE ------------------------------ closed 12078 Test timeout 60 s or wait for inversion of Acknowledge with inversion of LLS_SELFTESTBIT LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 76
FPA = 500 ml/min Valve Status: VEBK1/2, VABK1/2 --------------------- closed VDEBK1/2, VDABK1/2 ----------------- opened VBP, VDE ----------------------------- opened VDA ----------------------------------- closed PDA ≤ +380 mmHg or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 77
Pressure reduction FPE = 500 ml/min FPA = 500 ml/min Valve Status: VEBK1/2, VDEBK1/2 ------------------- opened VABK1/2, VDABK1/2------------------- opened VBP, VDA, VDE ------------------------ opened Timeout 60 s 1299 Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
• UFP is started during the pressure build-up phase. • The acknowledgement of the single test steps in the DFS pressure test is performed via a bit in the internal communication (between LLS and LLC). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
12036 VDE, VBP ------------------------------ closed 12037 VDA ---------------------------------- opened Final Condition: 12038 Test timeout 30 s or wait for inversion of Acknowledge with inversion of LLS_SELFTESTBIT LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
12076 VBP ---------------------------------- opened 12077 VDA, VDE ------------------------------ closed Final Condition: 12078 Test timeout 60 s or wait for inversion of Acknowledge with inversion of LLS_SELFTESTBIT LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
FPA = 100 ml/min Valve Status: VEBK1/2, VDEBK1/2, VABK1/2 --------- opened VDABK1/2 ----------------------------- closed VBP, VDA, VDE ------------------------ opened Final Condition: PDA ≥ +425 mmHg for 1 s or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
∆PDA < 25 mmHg, 12131 VBP, VDE, VDA ------------------------ opened 12132 Final Condition: 12133 Acknowledge with inversion of LLS_SELFTESTBIT Test timeout 60 s or wait for inversion of LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
FPA = 500 ml/min Valve Status: VEBK1/2, VDEBK1/2, VDABK1/2 -------- opened VABK1/2 ------------------------------ closed VBP, VDA, VDE ------------------------ opened Final Condition: PDA ≥ +425 mmHg for 1 s or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
VEBK1/2, VABK1/2 --------------------- closed VDEBK1/2, VDABK1/2 ----------------- opened VBP, VDA, VDE ------------------------ opened Final Condition: Test timeout 60 s or no increase of the values of the membrane position sensors B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
∆PDA < 25 mmHg, 12181 VBP, VDE, VDA ------------------------ opened 12182 Final Condition: 12183 Acknowledge with inversion of LLS_SELFTESTBIT Test timeout 60 s or wait for inversion of LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
PDA < -100 mmHg, 15033 Test timeout 1 min or wait for switch-off VGD Test delivery rate OSP at 200 ±15 ml/min 15034 – from LLS Acknowledge with switch-off of VGD 15039 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 92
∆PDA < 50 mmHg, Test timeout 1 min or wait for switch-off VGD from LLS PDA < -200 mmHg, 15112 – Acknowledge with switch-off of VGD 15118 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
3 revolutions have been detected Acknowledge with switch-off of VGD 20006 – 20011 2020x Wait for D24OFF (timeout 10 s) VGD ON 20012 2030x Wait for D24ON (timeout 10 s) 2099x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
The heater is switched off at TSD-S > 41 °C Acknowledge with switch-off of VGD 22006 – 22011 2220x Wait for D24OFF (timeout 10 s) VGD ON 22012 2230x Wait for D24ON (timeout 10 s) 2299x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
±40 mmHg. (Error) the test prerequisites from HDFonline-Task are not given during one of the test steps (e.g. autopriming, HDF filter rinsing required after LF error). Completed Close VPD B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Wait for B24ON (timeout 15 s) 4070x (Error) the test prerequisites from HDFonline task are not given during one of the test steps (e.g. autopriming, HDF filter rinsing required after LF error). 4099x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
(Timeout 35 s since start of subcode 400) Close VBT Test VBT closed 33039 |PA - PV| ≥ 200 mmHg 33040 Acknowledge with inversion of LLS_SELFTESTBIT 3350x Wait for inversion of LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Wait for B24OFF (timeout 7.5 s) VGB ON 37010 3730x Wait for B24ON (timeout 5 s) 3780x (Error) at start of test one or both couplings connected to rinsing bridge. 3799x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Hard Disk Drive USB8 USB7 USB6 USB5 USB4 Floppy Disk Drive USB2 USB1 USB1 Service SW/FSU PS/2 COM2 COM3 COM4 COM1 USB3 (DIABUS) (ABPM) (Touch Screen) (FPB) Type A B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
3 - 40 SW 9.xx 3.4.4 Detection/Boot Sequence from Bootable USB Sticks Use only USB sticks supplied by B. Braun as tool for the installation of the software. Dialog+ machines detect automatically connected bootable USB sticks. This is indicated by the red Installation text in start screen.
3. Press the TAB key when the logo is displayed. Main Processor..4. Press the Pause/Break key to stop the boot routine. The BIOS ..version is displayed, e.g. BIOS Revision 1.04..Pause Break B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 104
[Normal High] SHFCLK Active Period [Free Running] LP Active Period [Free Running] Onboard Audio [Disabled] Onboard USB1.1 [Enabled] Onboard USB2,0 [Enabled] Onboard DIE [Enabled] Memory Hole At 15M-16M [Disabled] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 105
PNP OS Installed [No] Init Display First [PCI Slot] Reset Configuration Data [Disabled] Resources Controlled By [Auto(ESCD)] x IRQ Resources Press Enter x Memory Resources Press Enter PCI/VGA Palette Snoop [Disabled] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Open rear door. Switch to TSM service program mode (on digital board DB, service switch S1 position 2). Switch on Dialog+. Document default values in default table. Switch off Dialog+. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 109
Position 2: TSM Service Program Mode Supervisor LEDs 0 to 7 (D28 – D35): Position 3: Software Installation/Update Mode Status 0 – 7 for Installation of LLS Software Jumper JP2 SWP: Supervisor Write Protect B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 110
JP1 and JP2 have to be set to this position (Default CWE and SWE: controller/supervisor write enabled). Set the jumpers back to the CWP/SWP setting after installation. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 111
Bootable USB stick with the TLC/LLC/LLS software 14. Open front door. 15. Connect USB stick A to USB port B on the TLC sub-rack C. 16. Switch on machine. Fig.: Front View TLC Sub-Rack B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 112
USB Stick not Detected The system does not detect the bootable USB stick B BRAUN (indicated by the green logo in the start screen). If necessary disconnect the USB stick and reconnect. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 113
If YES, press Y or y and <CR> if keyboard is used otherwise press the EQ hardware key! If NO, press N or n and <CR> on keyboard or any other hardware key! Press to confirm the selected CFC. [EQ] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 114
Don't forget to set the operation mode switch to TSM mode (2) and check in TSM all subsystems' version number. Switch the power OFF and REMOVE the USB stick before starting the Dialog... The TLC software was installed successfully. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Enter Customer Specific Default Values The customer specific default values of the machine must be entered after the software installation. 27. Enter customer default values from default table and save B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
33. Set system configuration (customer specific values) or if available use the FSU USB stick with the system configuration. 34. Save settings. 35. Switch off machine. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
(The ABPM function (if present) can be checked without a test run of the machine in preparation mode, without the concentrate pump.) 12. Terminate dialysis. 13. Perform disinfection/decalcification. 14. Remove blood lines, filter and heparin syringe. 15. Switch off Dialog+. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
The field service utilities has the following directory structure (see left figure). BlueCat DISKTYPE USER 04_Write_Languages 05_Write_Calibration_Data 06_Write_System_Configuration 09_Card_01_Read_Card_Image 09_Card_02_Write_Card_Image 09_Card_03_Read_Patient_Diskette 09_Card_04_Write_Patient_Diskette_to_Card 09_Card_05_Read_Patient_Card 09_Card_06_Write_Patient_Card 10_Read_System_Configuration 11_Read_Calibration_Data 12_Read_All_Trends 13_Read_Screenshots 14_Get_Error B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
To prevent the corruption of USB stick, before removing it, please press AQ, then follow the instructions on the screen! 4. Press the EQ hardware key to execute the utility. Press the AQ to exit the FSU program. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Card Commands in the main menu. 2. Press the EQ hardware key to execute the utility. 3. Press the BP hardware key and move the cursor down to select a utility. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Press the AQ hardware key to exit the FSU menu or to skip back to the previous menu. Press the EQ hardware key to execute a selected utility. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
BP stop stop hardware keys. Note: (n.nn = the version number) 4. Press EQ to accept the entered value. The main menu is displayed after a few seconds. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
3. Set the new working time counter with the BP and BP hardware keys. stop stop Press EQ to accept the entered value. The main menu is displayed after a few seconds. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
A r e y o u s u r e ? 3. Press AQ if you do not want to delete the masterboot record. AQ: no EQ: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 125
A r e y o u R E A L L Y s u r e ? the masterboot record. 5. Press BP to cancel the operation. BP-: no BP+: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
The last write procedure can be undone with the 07 Undo Last Written utility. The Dialog+ can be switched off and on again between the write procedure and using the 07 Undo Last Written utility. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 127
After successfully copying the language to the CFC the version numbers of all languages available on the Dialog+ are displayed. The language selected for treatment is also marked. 4. Press EQ to continue. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 129
(SYSver) stored in the calibration data file with the serial number and the system version number from the Dialog+. The calibration data file can be written only if they match. Example Ser-No.= 90832 SYSver = V8_20 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 130
The pairs that do not match are displayed with the following message: (not selectable) 5. Press EQ if you are sure. This will write the calibration data from the USB stick to the Dialog+ hard disk drive (CFC). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 132
Yes, system configuration file is written to Writing System Configuration to Dialog+ from: TLCver=90832/VX_XX Dialog+ hard disk drive (CFC) A r e y o u s u r e ? AQ: no EQ: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 133
The pairs that do not match are displayed with the following message: (not selectable) 5. Press EQ if you are sure. This will write the system configuration data from the USB stick to the Dialog+ hard disk drive (CFC). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
3. Press EQ if you are sure. This will undo the last written utility from the USB stick to the Dialog+ hard disk drive (CFC). AQ: no EQ: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Searching for md5sum files....the main menu after the self test. MD5SUMS.XXU.Win found..MD5SUMS.XXU.Lin found..MD5SUMS.SPEC.Lin found..Checking file system integrity..O.K. Calculated CRC = XXXX Reference CRC = XXXX B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
05. Read Patient Card (USB Stick → Card) 06. Write Patient Card 07. Remove All Patient Data BP-: move cursor up BP+: move cursor down AQ: back to previous menu execute B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 144
U S E R / 0 9 _ C a r d _ 0 3 _ R e a d _ P a t i e n t _ D i s k e t t e ....E n d o f Remove All Patient Data . . . B. Braun Avitum AG BA-TE-DE08C M.KAY...
BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 146
....displayed....End of Read System Configuration... B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 148
....... . End of Read Calibration Data... B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Year 2007 09_Card_03_Read_Patient_Diskette M M : 09 month September D D : 12 Wednesday 09_Card_04_Write_Patient_Diskette_to_Card h h : 14 hours 09_Card_05_Read_Patient_Card m m : 25 minutes 09_Card_06_Write_Patient_Card 10_Read_System_Configuration 11_Read_Calibration_Data 12_Read_All_Trends B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 150
....... . End of Read Trends... B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Dialog+. The vga*.zip files copied to USB stick are deleted from the Dialog+ hard disk drive (CFC) (similarly to the log files) but the subdirectory structure on the Dialog+ hard disk drive (CFC) is left untouched. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 152
....... . End of Read Screenshots... B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
3 - 91 SW 9.xx 3.6.10.5 Get Error (ErrorDisk) The error files can only be analysed by qualified personnel of B. Braun Avitum (e.g. development department) and should be provided only on special request. The utility gets error information and log files from Dialog+ hard disk drive (CFC) and writes them to the USB stick.
Page 154
Dialog+ hard disk drive (CFC) to the USB stick... The Get Error (ErrorDisk) menu automatically returns (a ..few seconds) to the main menu End of Error Disk... is displayed..... End of Error Disk... B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
USER/XX_Read..USER/14_Get_Error These files can be copied to any location on the PC (e.g. to use them for local examination) or can be deleted from the FSU USB stick. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
FSU program. - - - S W I T C H O F F t h e D i a l o g + - - - B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 157
Extended TMP Limit Range Button is Displayed in Treatment Select Extended TMP Limit Range inactive inactive inactive Arterial Bolus Parameters Arterial Bolus Volume Arterial Bolus with SAKA Support B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 158
2000 2000 Therapy Time h:min 04:00 04:00 03:30 Gross UF Rate vs. Blood Flow inactive inactive inactive Gross UF Rate vs. Blood Flow active active active UF Profile Editor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 159
Default Setting for Treatment Mode: Warning of Kt/V Target-Projected Deviation Warning active Warning active Warning inactive Warning for Saving Kt/V Table Timings Parameter Setting Window Disappearing Time Screen Saver Appearance Appearing Time B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 161
HDF Online Substitution Pump OSP (2.15) Unit Default US Default Modified µl/Head Revolution 8300 8300 Level Reg ulatio n ( 2.16) Unit Default US Default Modified LRP (PPR) Slow LRP (PPR) Fast B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 162
Morning 06:30 Rinsing Tuesday Night Morning 06:30 Rinsing Wednesday Night Morning 06:30 Rinsing Thursday Night Morning 06:30 Rinsing Friday Night Morning 06:30 Rinsing Saturday Night Morning Sunday Night Morning B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 163
Temperature Conductivity Name Volume Long Short Time min. (max. 20 Characters) h:min h:min h:min mS/cm Citric Thermal 00:15 00:05 00:05 Decalcification Short 00:05 00:05 00:05 Bleach 00:15 00:05 00:25 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 164
Lo 15 xevonta Lo 18 xevonta Lo 20 xevonta Lo 23 xevonta Hi 10 xevonta Hi 12 xevonta Hi 15 xevonta Hi 18 xevonta Hi 20 xevonta Hi 23 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
LLC Calibrations Check and set ratio, tubing constants BPA/BPV, OSP tubing type, check/set PA/PV limit values. Working Time Counter • Set working time counter. Motherboard Check and set BIOS. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 167
5. Set the blood pump so that no blood side alarms are activated. 6. Start dialysis. 7. Check alarm function of SAD. (The ABPM function can be checked without a test run of the machine in preparation mode.) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 168
1. The measures were executed correctly. Fill in the execution protocol after service (e.g. exchange of a component) and after executing all necessary measures according to the repair matrix. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 169
Inspection Protocol for Automatic Blood Pressure Measurement ABPM} Name Service Technician: The respective measures were executed correctly after servicing according to the repair matrix................................Date / Signature B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd...
Page 171
Test 1.14 Bypass and Disinfection Valves 4-88 4.8.2.15 Test 1.15 Flow Pump FPA, Balance Chamber Sensors and Air 4-89 Separator 4.8.2.16 Test 1.16 Pressure Water Part 4-91 4.8.2.17 Test 1.17 UF Pump 4-92 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 172
4.8.3.10.3.1 Turbidity Calibration with Water 4-144 4.8.3.10.3.2 Calibration with Reference Solution (Red Dye) 4-144 4.8.3.10.4 Actual Value Check Blood Leak Detector 4-145 4.8.3.10.4.1 Check 0 ‰ Blood Concentration 4-145 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 173
4-170 Currentless Closed SAKV-SG 4.8.3.27 Setting Servomotor for Disinfection Valve VD 4-171 4-173 Production Report 4-174 4.9.1 Version Number Dialog 4-175 4.9.2 Low Level Options 4-176 4.10 Service Reports B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Position 3 = Software Installation/Update 4. Switch on machine. The machine starts in TSM mode. Online Board Power Board M otors Power Board Valves Digital Board P1 1 Basic Board Analog Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 175
The File Operations menu can be activated with this icon. The current software/firmware versions are displayed in the Dialog Version Number menu. This menu can also be selected in the Production Report. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
3. Close rear door. 4. Switch on machine. The machine starts in therapy mode. Online Board Power Board M otors Power Board Valves Digital Board P1 1 Basic Board Analog Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
7500 Malfunctions in the balance chamber can occur if: • the values are out of limits • the minimal difference from the detected min. value was not reached B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 179
L L C / S & T L C A l a r m W i n d o w L L C M e s s a g e W i n d o w Fig.: Second Page Service Overview in Therapy B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 180
L L C / S & T L C A l a r m W i n d o w L L C M e s s a g e W i n d o w Fig.: Third Page Service Overview in Therapy B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
LS CURRENTS_S: Loudspeaker current supervisor LS PULSE_S: Loudspeaker pulse supervisor MSBK1: Membrane position sensor balance chamber 1 MSBK2: Membrane position sensor balance chamber 2 NSVB: Level sensors upline tank B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 183
LLC text messages according to the LLC message hex numbers from LLC L L C M e s s a g e W i n d o w message window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
5500 Malfunctions in the balance chamber can occur if: • the values are out of limits • the minimal difference from the detected min. value was not reached B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
LLC/LLS after every calibration. The calibration data can be downloaded from the compact flash card CFC (hard Down Load Calibration Data disk drive) to the exchanged digital board DB. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 188
• Basic Board BB • Analog Board AB • Power Board Motors BPM • Power Board Valves BPV • HDF Online Board HOB • Digital Board DB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
The new settings are valid after the restart of the machine. There are five trend groups, which can be selected for the treatment program. Trends and Trend Groups Every group has three treatment parameters. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
1. Click the mmol/l or mS/cm key to select a measurement unit. The acetate mode can be disabled. Acetate Mode Disabled 1. Click the YES key to disable the acetate mode. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 191
[mS/cm per mmol/l]. Conversion Factor: Conductivity mS/cm (Standard Value) Conversion Factor NaCl Concentration mmol/l (Measured Value) 4. Click the Default Value window if necessary to change the value [mS/cm]. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 192
[mS/cm]. 4. Click the ACID icon if necessary to edit the following parameters: • Acid Concentrate Name • Conversion Factor [mS/cm per mmol/l] • Default Value B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 193
1. Click the ml/min key to set the dialysate flow. You can set the dialysate temperature for therapy. Dialysate Temperature 1. Click the C key to set the dialysate temperature. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
The selected Chamber-free AV Line does not appear. AV Line with PBE Connector Activate if AV lines with PBE pressure connection are used. Do not activate if AV lines without PBE pressure connection are used. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
You can set the upper relative alarm limit for the Delta PA max. PA limit window. Delta PA PA max. max. PA ref 200 mmHg PA min. Fig.: Limit Window for PA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 196
The TMP low limit range is set min. –10 mmHg. If the key is not pressed The TMP limit range is extended to –100 mmHg If the key is pressed (absolute TMP value). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 197
• a warning message is not displayed anymore TM P 20 mmHg 12 - 28 m m Hg • treatment is interrupted • machine is switched to bypass -100 Bottom TM P Lim it Value B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Arterial Bolus with SAKA Support Arterial Bolus Parameters menu. Prerequisites are: • Arterial tubing clamp SAKA present • An arterial line system with infusion access is present between the SAKA and blood pump B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Furthermore the control pressure PV can be set. Thereby the blood pump is started/stopped and the venous tubing clamp currentless closed SAKV-SG is opened/closed. The control pressures for PA and PV can be set for single-needle cross-over. Single-Needle Cross-Over B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Heparin Bolus Volume Therapy Beginning Bolus An automatic bolus can be administered at the beginning of a therapy. The heparin bolus is performed by the venous red detector RDV. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 201
You can set the syringe name and the inner diameter of the syringe. Heparin Pump Compact 3. Activate the Selected Syringe menu with the Selected Syringe icon. The following menu is opened. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
• Rinsing Blood Flow • Rinsing Dialysate Flow • Rinsing Time by UFP • Rinsing Rate by UFP • Rinsing Volume by UFP • Blood Flow for Connecting Patient B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 203
You can set the blood flow for connecting the patient (in therapy). The blood Blood Flow for Connecting Patient pump BPA is started with this rate when the therapy is started. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
The alarm can be acknowledged with the key. Setting Range The alarm limit can not be set lower than the warning limit (even if the warning limit is deactivated). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 205
A linear adaptation is calculated for the remaining process. EXP (Exponential) If this key is activated only the first interval can be set. An exponential adaptation is calculated for the remaining process. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 206
The UF process of the activated interval is set to HD with the HD key. UF Mode Sequential SEQ The UF process of the activated interval is set to sequential with the SEQ key. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
HDF Dialysate Flow for Predilution HDF/HF Infusion Bolus Volume The HDF/HF infusion bolus volume can be set. You can set predilution for a HDF/HF online therapy by activating the Post/Predilution Selection Predilution key. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
• Auto Switch-On without Daily Confirmation • Maximum Out of Action Time (Day/Hour/ Minutes) • Disinfection Configuration Data Chemical (long), Chemical (short) • Disinfection Configuration Data Central Thermal, Central Chemical, Rinsing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 209
The machine is rinsed after the preselected switch-on time. The key is only enabled if the function is selected. Switch-On Time The switch-on time can be set (without a weekly schedule). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 210
The function is activated if the Limit OFF key is not pressed. The Limit OFF key is activated in the default setting, the parameter keys are disabled, i.e. parameter values can not be set. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 211
• Rinsing Time Central Automatic Chemical: • Inlet Volume • Switch-off for Retention without Automatic Switch-on • Retention Time • Rinsing Flow • Rinsing Time Rinsing: • Inlet Flow • Time B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 212
FBK1 VEBK1 VDABK1 M SBK1 VDEBK2 VABK2 BICP LAFS VEBK2 VDABK2 FBK2 M SBK2 RVDA TSHE RVFPA RVUFP Fig.: Flow Diagram - Central Automatic Chemical Disinfection of Water Inlet B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 213
The rinsing flow can be selected. Rinsing Time The rinsing time can be selected. Central Chemical A central chemical mode can be selected for the automatic switch-on in the weekly schedule. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
The operating hours can be set for the intended operation time of the DF/HDF Filter Operation Time [h] filters. The operating hour counter is active when DF Preparation is active. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
V (litres)=2.447 – 0.09516 x Age (years) + 0.1074 x Height (cm or feet/inch) + 0.3362 x Dry Weight (kg) Female: V (litres)=-2.097 + 0.1069 x Height (cm or feet/inch) + 0.2466 x Dry Weight (kg) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 216
The Kt/V is approximately calculated via the Dialyser Filter Urea Clearance Data for the selected dialyser (see table, use scroll bar to see table). Details can be found in Chapter 3 repair instructions. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Default Setting for Treatment Mode: value, a message is activated. Warning of Kt/V Target-Projected Deviation A warning message can be enabled for saving the Kt/V table. Warning for Saving Kt/V Table B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Therapy and end of therapy: Time cake, remaining time, status line Disinfection: Time cake, disinfection mode, phase You can set the appearance time of the Screen Saver. Appearing Time B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
English. If German is selected the therapy and TSM language is German. If an additional language is selected this language is used in therapy. The TSM language is English. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
The limit values in therapy, e.g. for systole/diastole can be stored on a patient Taking Over Limits from Patient’s chip card. These limit values are taken over if the patient chip card is read. Parameters B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
The ID number 200 indicates that the Guide Line Mode is enabled and can be used in therapy. ID Number 100 The ID number 100 indicates that the Guide Line Mode is not enabled and can not be used in therapy. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
If necessary you can set the time difference between summer and winter time. Time Difference Note: The Time Difference should be set to 00:00 for countries where daylight saving time is not applicable. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
It is prohibited to run a therapy with a patient if the Skip Self Test function is activated. The self test must be passed successfully prior to a therapy. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 224
A window is automatically opened at the end of therapy. This window must be Automatic Reinfusion Start at confirmed at the end of the therapy for the reinfusion of the blood. Entering End of Treatment B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Nexadia functionality, e.g. support of the option bioLogic RR Comfort and Adimea; the mmol/l unit can be used. The Nexadia protocol must be supported by the host computer. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
9 to 18 month period if the 4,000 working time is more than 4,000 hours after the last reset. W indow 4,000 hours vs 9 to 18 months Elapsed Months B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 227
Press the Preventive Maintenance Accomplished key and then the Preventive Maintenance Accomplished Save icon after the preventive maintenance was performed. Consequently a new preventive service period is started again after the machine is switched off and on. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Trend and Trend Groups key in the Treatment Support menu. The following menu is opened and displayed. 3. Activate the Trend History menu with the Trend History window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Page 229
5. Activate the Edit menu with the Edit icon. 6. The treatment parameters in the six trend groups can be modified. 7. The default parameters can be reset with the Set Defaults window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd...
Check red LED for signal lamp (on optical status display boards OSD). Check yellow LED for signal lamp (on optical status display boards OSD). Check green LED for signal lamp (on optical status display boards OSD). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
1. Activate the Monitor Test menu with the Monitor Test key in the TL Manual Test and Calibration menu. The following menu is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
(on the keyboard membrane of the TFT housing) three times in the TSM Main Menu. 2. Switch machine off and on again. The old default setting is loaded. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The day brightness (digits) for the backlight set. Test OSD Red and Yellow (for Signal Lamps) The red and yellow LED of the optical status display OSD can be checked, i.e. switched ON and OFF. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
• Use an empty service chip card (art. no. 7703856) for the test and not a patient chip card. • Do not remove the chip card from the card reader during the test. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The maximum speed of the USB port. Status The connected USB device is in the valid USB device list. Not OK The connected USB device is not in the valid USB device list. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Parameter Check key. The following menu is displayed. 3. Press the key to close the supervisor window. Press the key to select and acknowledge the single parameters and close the supervisor window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Hard Disk Check (for the compact flash card) to test the hard disk drive. Press the key of the Application SW Check to test the application software. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
ABPM system is tested. The measurement chamber is connected to the ABPM system via the pressure tubing. A pressure is built up to 300 mmHg and is held for 3 minutes. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
5. Switch off machine after successful test. 6. Remove the test plug from the RS 232 port. Pin Assignment RS 232 Test Plug (Soldering Side) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
(residual digits). This value must be ≤4000 digits (maximum service life). Example: Required current: 1332 (digit) Reserve: 2668 (digit) Total digits: 4000 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
If the Sent Data and Received Data are identical, the test was successful. 6. Switch off machine after successful test. 7. Remove the test plug from the DSI interface. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Previous and Next Arrow keys. If required, a corresponding calibration menu can be selected in certain LL Manual Test menus. The calibration procedures are described in paragraph 4.8.3 Low Level Manual Calibration. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
• -12 VAN > -11 V -12 VAN Control / Supervisor ON Test activated ±12 VAN Test: • +12 VAN < 11 V • -12 VAN < -13 V B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The motor torque is displayed by the controller. Torque Cover State If the pump cover is opened the blood pump BPA is stopped. The controller detects and displays the opened pump cover. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The motor torque is displayed by the controller. Torque Cover State If the pump cover is opened the blood pump BPV/substitution pump is stopped immediately. The controller detects and displays the opened pump cover. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
(The clasp nut is assembled on the drive spindle and limits the max. pressure load of the heparin pump). Activates/switches off the heparin pump. HP Motor State Selection between slow and fast forward feed. HP Speed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Thereby the switch function and leakage test (pressure - clamp SAKV and SAKA gap, time) can be tested. The closed clamp gap is set to 1.4 mm + 0.1 mm. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Machine with Single Pump The value for PBS is 0. Machine with Double Double Pump must be activated in Production Report, Low Level Options. A value for PBS is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
A tubing filled with fluid is inserted in the SAD If the SAD TEST is activated NO AIR must be detected. The light path is interrupted NO BLOOD (Controller/Supervisor) changes to BLOOD (Controller/Supervisor). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
You can check the actual status of the reed sensors. Detectors Bicarbonate rinsing connection sensor. BICSS Concentrate rinsing connector sensor. Rinsing bridge connector sensor (dialyser coupling). SBS1 SBS2 Rinsing bridge connector sensor (dialyser coupling). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
All balance chamber valves are opened and the outlet flow pump FPA works in main flow with the same speed as the inlet flow pump FPE. The degassing pump EP runs and the degassing bypass valve VEB is open. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
C or the upline tank VB is empty. The degassing pump EP must be able to build up a flow of > 1200 ml/min at a negative pressure of - 500 mmHg. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Increase the speed [rpm] until the flow pump FPE has built up a flow of 750 ml/min. The measurement range is as follows: TSE, TSD, TSBIC: 0 to 100 TSD-S: 30 to 55 TSDE: 30 to 55 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Increase the speed [rpm] until the flow pump FPE has built up a flow of 750 ml/min. Thereby the pulsating flow of the pumps have no influence on the conductivity measurement. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The following valves can be activated: Valves Bypass valve Dialyser inlet valve Dialyser outlet valve Disinfection valve Circulation valve Air separator valve B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
FPE. Simultaneously the fluid in the right side of the balance chamber is emptied to the drain. Drain Fig. : Filling Cycle Balance Chamber Phase 1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Page 259
Thereby FPE delivers fluid into the Fig. : Filling Cycle Balance Chamber Phase 2 dialyser circuit. VLA is only closed and switched to normal operation if the top level sensor LAFSO detects fluid. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The pressure can be displayed in mmHg or in mbar, depending on selection. Conversion Factor mmHg to mbar Conversion Factor = 1.3332 Example: -100 mmHg approx. –133 mbar –400 mmHg approx. –533 mbar B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
UFP, set by the value of the throttle DDE. Not applicable. Flow Detector Stroke Value UFP The Stroke Value UFP should not be used in a normal case, because UFP must be calibrated (see LL Manual calibration). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
If air is visible check the complete system for leaks in menu 1.20 Test Water Part. After the initialisation phase rinse the blood leak detector approx. 1 min and then perform a leakage test of the system in menu 1.19 Leak Test. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Page 263
LAMP TEST is switched back to OFF. The calculated value must be in the following limits (to the first decimal digit/round off): red LED / green LED 1.15 - 1.44 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The test is started with a negative test pressure and then followed by a positive test pressure. The leakage test consists of three test steps. If an error occurs an LLC alarm is LLC Alarm activated (LLC Status 00020000). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Page 265
VDEBK1 VABK1 FBK1 Z.D. VEBK1 VDABK1 M SBK1 Dialyser VDEBK2 VABK2 V.D. LAFS VEBK2 VDABK2 FBK2 M SBK2 SBS1 RVUFP RVFPA SBS2 Fig.: Excerpt Flow Diagram for HDF Online B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Page 266
Rinse water circuit to prevent air bubbles. b) Close the outlet. c) Activate test. The FPE builds up a pressure. d) Check pressure drop. The pressure drop must be < 50 mmHg in 30 s. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Valves VDA/VDE are open (valve symbols are black). Valve VBP is closed (valve symbol is white). Bypass Valves VDA/VDE are closed (valve symbols are white). Valve VBP is open (valve symbol is black). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The actual value, i.e. the fluid level of the upline tank VB is displayed by the level sensors NSVB. Start the Emptying ON at least three times and manually clamp the bypass (tubing) to ensure a sufficient emptying (for transportation). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 - 100 SW 9.xx 4.8.2.22 Test 1.22 HDF Not applicable. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Valves VVB, VBICP, VBKO, VBKS The valves VVB, VBICP, VBKO and VBKS can be switched either separately or together. The status of the reed sensor (OPEN/CLOSED) is displayed. Detectors BKUS, BKUS_S B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
DF filter and HDF filter are impermeable to air. Valve VBE can only be opened if PDA < -100 mmHg to prevent fluid penetration. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The status of the additional coupling sensor FEDFFS can be checked. FEHDFS (filter recognition HDF filter) The status of the additional coupling sensor FEHDFS can be checked. FEHDFS PSAUS PSABFS FEDFFS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Akku OK (Battery OK) time Battery capacity < required capacity for 20 minutes battery running time External State Indicates if machine was switched on externally. Machine was switched on externally B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The test codes for LLC and LLS are displayed for all test phases. In case of an Test Code error the defective components can be determined with the help of the test description. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
FPB and wait for switching (OFF/ON). • Enter Key Press the key on the front panel board FPB and wait for switching (OFF/ON). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
C and is then switched off. The blood pump can be activated additionally if the measurement is performed with tubing set (line). Main Flow The machine can be switched between main flow and bypass. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
> 30 s (disconnect mains plug from mains receptacle). #115 Buzzer Test by LLS Ver/SW The software version of the SMPS-MC is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
• LLC sends #115 to SMPS-MC • LLS must activate AKAL-S within 5 s • Buzzer OK. if not OK: not OK: • Warning is generated • Buzzer not OK. • Therapy still possible B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
The delivery of the LRP can be checked with the Test Test key. A measurement chamber must be connected to PV. LRP stops when 500 ml are reached and the required time for the pressure build-up is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd...
Input values can be increased in increments with the displayed hardware key as an alternative to the software key on the touch screen. The service overview can be selected in the TSM with the hardware key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 281
If required, the previous calibration menu or the next calibration menu can be selected with the Previous and Next Arrow keys. The menu can be closed with the Exit Door key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 282
C o n t r o l l e r / S u p e r v i s o r ( 1 2 b i t ) Voltage 0 to 5 V Digits 0 to 4095 B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
LL Manual Calibration key in the Manual Test and Calibration menu. Activate the Pressure Sensors Blood Side menu by pressing on the line Calibration 2.1 Pressure Sensors Blood Side. The following menu is displayed. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 285
Build up a pressure of -100 mmHg (+10 mmHg) with a syringe and clamp the tube. -100 mmHg Press key. Wait until digits are stable. Enter the stable value from the pressure measurement instrument as reference value. Confirm with the key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 286
Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 16. Confirm and store completed calibration with key. 17. Remove pressure measurement instrument, syringe and tubing. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 - 118 SW 9.xx 4.8.3.4 Heparin Pump Not applicable. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
You can calibrate the degassing pressure sensor PE. Build up the negative pressure in the degassing circuit with a syringe. RVDA SKBS VVBE DBK* TSBIC VB WT VBICP* VBKO* TSHE VVB* VBKS RVBO* Fig.: Degassing Circuit and PE B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 289
TSHE Fig.: Water Sub-Rack with Loop-In for Manometer at PE and Tubing Fig.: Water Sub-Rack with Tubing Clamp between FEP and EP (front Clamp at VEB (rear view) view) B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 290
Manual Test and Calibration menu. 2. Activate the Pressure Sensors (Water Side) menu by pressing on the line Calibration 2.3 Pressure Sensors (Water Side). The following menu is displayed. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 291
PE should show a of approx. < -35 mmHg/-50 mbar • Close VEB and increase speed to 2,100 rpm • PE should show a value of approx. < -500 mmHg/-670 mbar B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
15. Confirm and store complete calibration procedure with key. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 16. Remove calibration devices. 17. Reconnect tubings. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
VDA, it must be closed or the tubing must be clamped manually. FBK2 TSDE DF Filter PEV* Z.D. Dialyser V.D. LAFS FBK2 BL PDA SBS1 RVFPA SBS2 Fig.: Dialysate Circuit and PDA B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 295
LL Manual Calibration key in the Manual Test and Calibration menu. 2. Activate the Pressure Sensors Water Side menu by pressing on the line Calibration 2.3 Pressure Sensors Water Side. The following menu is displayed. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 296
21. Confirm and store complete calibration procedure with key. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 22. Remove calibration devices. 23. Reconnect tubings. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
KP/BICP TSHE Keep tubings as short as possible TSHE RVDA RVDA Fig.: Original Tubing Water Sub-Rack for TSE/TSHE Fig.: Water Sub-Rack with TSHE in Line with TSE for Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 298
C) is displayed at the temperature measurement instrument. Wait until temperature is stable, i.e. 0.1 C for at least 10 s. Press key. 10. Enter the stable value from the temperature measurement instrument as reference value. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 299
22. Stop degassing pump EP and flow pump FPE, and disconnect adapter set. 23. Reassemble the heater inlet temperature sensor TSHE to the original position (see previous page). Note: Pay attention to correct flow direction of temperature sensor TSHE. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Measurement Instrument ml/min RVFPE ENDLF-S TSBIC blue mS/cm BICLF ENDLF TSD-S TSDE Measurement Sensor BICP blue FBK1 VEBK1/2 Fig.: Dialysate Preparation with Loop-In Point for TSBIC/TSD/TSD-S and TSDE Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 301
FBK1 Measurement Sensor blue TSDE Fig.: Tubing Rinsing Bridge with Removed TSDE Fig.: UF Sub-Rack with TSDE in Line with TSBIC/TSD/TSD-S (on DF Sub-Rack) for Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 302
Increase slowly the speed of FPE, until a flow of 750 ml/min ±50 ml/min is reached. Press key for a combined calibration of TSBIC, TSD/TSD-S and TSDE (The circulation valve VZ stays closed). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 303
4.8.3.1 General Calibration Information). 21. Reassemble the heater inlet temperature sensor TSDE to the original position (see previous page). Note: Pay attention to correct flow direction of temperature sensor TSDE. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
If acid concentrates are not available the following B. Braun acetate concentrates can be used as an alternative: e.g. SW44 and SW 174 If acid concentrate from B. Braun is not available acid concentrate from a second source can be used. The composition is listed in table 1. Composition of Calibration Acid concentrate is diluted with deionised water with a ratio of 1:4 (or 1:4.5...
Page 305
FBK1 Measurement VABK2 Sensor VDEBK2 FBK1 blue VEBK2 Fig.: Original Tubing UF Sub-Rack with FBK1 Fig.: UF Sub-Rack with Loop-In Point for BICLF, ENDLF and ENDLF-S Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 306
Connect measurement instrument between FPE and balance chamber. Connect dialyser couplings to flow meter. Increase slowly the speed of FPE, until a flow of 750 ml/min ±50 ml/min is reached. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 307
14. Press key. Wait until digits are stable. 15. Enter the stable value from the measurement instrument as reference value. 16. Confirm with key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 308
Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 27. Stop FPE and BICP and remove measurement instrument. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
SBS2 blue Fig.: Rinsing Bridge with Blood Leak Detector BL TSDE LAFS FBK2 SBS1 blue SBS2 Fig.: Rinsing Bridge with Red Coupling Immerged in Reference Solution for BL Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 310
Part. After the initialisation phase rinse the blood leak detector approx. 1 min and then perform a leakage test of the system in menu 1.19 Leak Test. Chemical Disinfection with Citric Acid 50% A decalcification must be performed before initial calibration. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Please Note The same blood leak detector is used for both settings: • The calibration procedure is identical, but with different tolerance ranges • The test procedure is identical B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
LAMP TEST is switched back to OFF. The calculated value must be in the following limits (to the first decimal digit/round off): red LED / green LED 1.15 - 1.44 B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
The displayed actual blood concentration value must be in the following limit, to ensure a sufficient resolution: Actual Status Controller > 200 ‰ Blood Concentration [ Confirm and store completed calibration procedure with key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
The blood concentration values for the controller and supervisor must be in the following limits: Actual Status Actual Status Controller Supervisor Blood Concentration 0.90 to 1.1 0.90 to 1.1 [‰] B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
• Balance or graduated cylinder • Tubing clamp Fill graduated cylinder with 200 ml water. Remove respective concentrate suction rod from rinsing chamber and immerge into graduated cylinder. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
12. Confirm and store completed calibration procedure with key. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
It is recommended to calibrate the UFP with the procedure described (with balance) because the accuracy of the stroke value might be insufficient for a proper UF control of the system. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
The calibration procedure is completed. 5. Press the key to store the calibration. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information 4.8.3.1 General Calibration Information). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Remove the tubing from the SAD. Interrupt the light path in the SAD with a piece of paper. The following settings must be displayed: RDV: NO BLOOD (Controller/Supervisor) changes to RDV: BLOOD (Controller/Supervisor) B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Production Report, LLC Options, Holder for BIC Cartridge. You can determine the limit of the ENDLF ratio (measured at the end conductivity sensor ENDLF). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 321
Connect the respective concentrate (bicarbonate/acid) after the request. Select service overview to display the current RATIO values. Wait till stable ratio values are reached. Use the mean value if there are slight deviations. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 322
1. Enter 35 in Desired Setting Controller for the BICLF Cart Ratio. 2. Save the entered value with the Calibration key. 3. Check the bicarbonate ratio value in therapy. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 - 154 SW 9.xx 4.8.3.14 Calibration Load Cell Not applicable. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
The supervisor uses a tolerance of 10 mmHg for monitoring the PV limit values to prevent double alarms. The set PA Absolute Low Limit value (-200 mmHg) is used as limit in therapy PA Absolute Low Limit and TSM service program/system configuration. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Pump Head Selection. The following menu is displayed. You can select the pump head roller type used in the blood pumps. • Pump head 8 x 12 • Pump head 7 x 10 (option) B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 - 157 SW 9.xx 4.8.3.17 Infusion Density Substitution Pump Not applicable. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
2.15 HDF Online Substitution Pump OSP. The following menu is displayed. The stroke value must be set to 8300 (for pump head 8 x 12). Stroke Value HDF Online Substitution Pump OSP B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
The settings are (see default table in chapter 3): • 60 % for the slow mode • 100 % for the fast mode The default settings should not be changed, unless instructed otherwise. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
You can calibrate the inlet flow pump nonreturn valve RVFPE. Fig.: Calibration RVFPE/Loop-In Point Loop-In Point for Calibration FBK1 Fig.: UF Sub-Rack with FBK1 RVFPE Fig.: DF Sub-Rack with RVFPE B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 330
Set a pressure of 975 mmHg (1300 mbar) with the adjustment screw of RVFPE. • 975 mmHg ±37 mmHg • 1300 mbar ±50 mbar Tighten lock nut again. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 331
VDABK1 MSBK1 VDEBK2 VABK2 VEBK2 VDABK2 FBK2 MSBK2 RVDA Fig.: Calibration RVDA/Loop-In Point Loop-In Point for Calibration FBK1 Fig.: UF Sub-Rack with FBK1 RVDA Fig.: Water Sub-Rack with RVDA B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Set a pressure of 975 mmHg (1300 mbar) with the adjustment screw of RVFPA. • 975 mmHg ±37 mmHg • 1300 mbar ±50 mbar Tighten lock nut again. 10. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
(30 to 40 °C). Thus uncontrolled pressure conditions are prevented. Select Test 1.20 Water Part (Overview) in TSM. Increase slowly the speed of FPA, until a flow of 500 ml/min is reached. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 335
Set a pressure of 300 mmHg/400 mbar with the adjustment screw of DDE. • 300 mmHg ±37 mmHg • 400 mbar ±50 mbar Tighten lock nut again. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
You can calibrate the pressure reducer valve DMV. RVDA SKBS VVBE DBK* TSBIC VB WT VBICP* TSHE VBKO* VVB* VBKS RVBO* Fig.: Calibration DMV VVBE Locking Screw for Manometer Fig.: Water Sub-Rack with DMV B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 337
0.9 ±0.1 bar with the adjustment knob. Close water tap. Press down the adjustment knob to lock pressure reducer valve DMV. 10. Switch off Dialog. 11. Disassemble manometer. 12. Reassemble locking screw. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 338
Note The 1.4 mm template fits through the gap. Check wit the 1.5 mm template. Note The 1.5 mm template must not fit through the gap. Reassemble cap C. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
SAKV-SG. • O-ring 7 x 1.5 art. no. 34570624 Check the o-rings C on the V-stopper/stopper D Replace if o-rings show any damages (tears/ruptures/ brittleness). Close front door. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Do not pull off the bow G from the control disc H, because its position is set ex works. The bow G on the control disc H must be seated in the groove F of the eccentric axle. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 341
D Light barrier E Bevel on eccentric axle F groove G Bow (only partly visible, is seated in the eccentric groove) H Control disc Rinsing bridge Draw-Up of Disinfectant B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
• The activation of the HCT Crit-Line option enables a window in therapy. This window allows the data exchange and display of measurement values to a connected HemaMetrics™ Crit-Line device. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 343
• Basic Board BB • Analog Board AB • Power Board Motors BPM • Power Board Valves BPV • HDF Online Board HOB • Digital Board DB B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
If the option is assembled or the option is retrofitted, it must be selected and WAN = Wide Area Network saved to activate the option. BSL = Bedside Link DBI = Dialog Bedside Link Interface B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
The Operation Mode Report can be selected in the Service Reports menu. Operation Mode Report The switching from therapy mode to TSM mode and back to the therapy mode is documented with date and time. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd...
Page 346
5. Technical Safety Inspection and Preventive Maintenance 1 / 2 0 1 0 5 - 1 SW 9.xx This document is the property of B. Braun Avitum AG with all rights reserved. Copyright Only trained personnel must service the Dialog+, i.e. repair, maintenance, software installation, Service firmware update, retrofitting and commissioning of the Dialog+.
Page 347
Measurement Circuits for the Measurement of the Electrical Safety According to EN 62353/60601-1 Inspection Protocol for Automatic Blood Pressure Measurement ABPM together with the Dialog+ 5.2.1 Execution Procedure for Measurement Inspection ABPM B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd...
Page 348
.................................. Operating Hours: ........h Inventory No.: .............. SW Version: ..........Manufacturer: B. Braun Avitum AG 34209 Melsungen, Germany Check List (S = Technical Safety Inspection Points; M = Preventive Maintenance Points) NO YES OK Note: Text in { } brackets is information for the execution of the check list! Technical Safety Inspection ...
Page 350
{The machine must be in therapy mode. Disconnect mains plug: The buzzer from the SMPS-MC gives three signalling tones as an indication that the machine is running in battery mode.} M 5.10 Disinfection: - Perform by responsible organisation {user} {inform responsible organisation} - Performed by service technician B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd...
Page 351
As a supplement this kit includes all parts recommended for exchange during a {34560686 Maintenance Kit/Supplement HDF Online} preventive maintenance of a HDF online machine.} B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd...
Page 352
The test current must be measured in both directions. Fig. 1: Protective Earth Resistance Equipment Leakage Current: • Differential Measurement L(N) N(L) Fig. 2: Equipment Leakage Current Patient Leakage Current Fig. 3: Patient Leakage Current B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd...
Page 353
Dialysis Machine Dialog REF {Type/Typ}: SN {Serien-No./Nr.} ..................ABPM Module M2009/2010 Automatic Blood Pressure Measurement ABPM B. Braun Avitum AG, 34209 Melsungen, Germany Manufacturer Address: Responsible Organisation (User) ......................................................................................................Visual Inspection - Tight seat of all connectors : ----------------------------------------------- OK - No damages which could affect the function or the safety (incl.
Page 354
Fig.: Inspection Set-up for Compliance of Error Limits of the 6. Decrease pressure according to test points and enter values in Measurement Display table. 7. Enter deviations in table. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd...
Page 355
: ---- max. 10 s Pressure reduction from >260 mmHg to <15 mmHg Pressure Tube 50 ml 50 ml Syringe Stopcock System Fig.: Inspection Set-up for Rapid Exhaust B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd...
Cable connection from basic board to front door: - Double pump - HDF online Cable connection from basic board to front door: - Double pump - HDF online B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd...
6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s 1 / 2 0 1 0 SW 9.xx 6.2.2 Wiring Diagram Front Door Single Pump Motor Heparin pump DZ/DR, Cover SAKV B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd...
6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s 1 / 2 0 1 0 SW 9.xx 6.2.3 Wiring Diagram Front Door Double Pump Motor Heparin pump PBE, PBS DZ/DR, Cover SAKV SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd...
6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s 1 / 2 0 1 0 6-10 SW 9.xx 6.2.4 Wiring Diagram Front Door HDF Online Heparin pump Motor PBE, PBS SAKV SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd...
Page 366
6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s SW9.xx 6.2.5 Wiring Diagram Housing B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd...
TSHE NSVB Water Block Option BIC VVBE VBKS VBKO VBICP bk black bl blue br brown gn green gr grey rd red vi violet wh white ye yellow B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd...
Silicone Silicone from Tubing Tubing WA Block 2.9x3 to BIC Suction Rod T Reduction on UF Sub-Rack to Concentrate to LA (UF) Suction Rod to TSHE/VEB (WA Block) B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd...
Silicone Tubing NSVB Water Water Water Inlet Inlet Inlet from Piston Pump Rinsing WA-Block from VLA Silicone to DF Tubing Processing Silicone Tubing SNP4 TSHE RVDA Silicone Tubing B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd...
Page 375
(Material Return Form) to B. Braun Avitum AG via the representative of the respective country for the ability to supply refurbished spare parts. The condition and the version of the defective spare parts must allow a qualified refurbishment.
Page 376
..................................Disposal and Taking Back of Spare Parts Dispose spare parts (e.g. boards or batteries) according to local disposal guidelines or send back to B. Braun Avitum AG free of charge. Cleaning and Disinfection of Spare Parts Used in Dialog Machines All spare parts sent by the customer to the B.
Page 377
Switch Mode Power Supply Microcontroller SMPS-MC 7-29 7.7.3.1 Compatibility Table SMPS-MC 7-30 Front Door 7-31 7.8.1 Manometer Connection 7-33 7.8.2 Heparin Pump 7-33 7.8.3 Blood Pump 7-34 7.8.4 Safety Air Detector SAD 7-35 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
7.6 for single parts UF sub-rack see paragraph 7.5 for single parts DF sub-rack see paragraph 7.4 for single parts Water sub-rack see paragraph 7.3 for single parts B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
C) without loosening the self-locking nut A. The procedure is only permissible if the coil can be turned without much iv 1 x Screw 40x10 effort. v 1 x Washer 4.2 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 382
TSHE 8 34570730 1 Housing complete for temperature sensor 9 3451727B 1 Throttle 0.4 bar 10 3451664B für HDF-Online 1 Throttle 1.3 bar 11 34517782 5 Disposable filter RVDA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 Motor for gear pump (SW 1.7.4) Micropump Version 2 for EP only 1b 34560319 1 Gear pump (Micropump version 2) (see A in figure) with magnet, without motor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
7.4.1 - DF block see separate figure for single parts, paragraph 7.4.1 see separate figure for - Piston pumps single parts, paragraph 7.4.2 BICLF BICP ENDLF TSDS RVFPE B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 Motor for gear pump (SW 1.7.4) 1b 34515330 1 Gear pump (Micropump version 1) Micropump Version 2 for FPA/FPE only (see B in figure) with magnet, without motor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 Motor for piston pump 5 34560572 1 Bell joint for piston pump 6 34560602 1 Pump body for piston pump - 7703848 required for motor assembly 1 Assembly jig (motor piston pump) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
10 34561277 1 Temperature switch assemble temperature switch on TL sub-rack Basic Board with heat sink paste Analog Board Online Board Motherboard Power Board Valves Online Board Basic Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 Compact flash card 1 GB The SW for the machine has to be installed again after replacing the CFC. 2 34560610 1 LX800 Motherboard 3 34519327 1 Battery BR2032, 3V Motherboard B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 Fan 92 x 92 mm with connector 2 34570284 5 Dust filter 3 3451059A 1 Mains switch 3 34770003 5 Protection cap for mains switch - Mains cord see separate figure (paragraph 7.7.1) SMPS-MC B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 397
4 34518029 1 Strain relief 5 3451441A 1 Tubing holder/fan cover compatible to crank 5a/5b 5a 34518355 1 Crank for blood pump 5b 34560157 1 Crank for blood pump B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Central Earth F5B F5A Cable Gland with Lock Nut Rear Door Cable Ties N PE PE LH1LH2 Ferrite Ferrite Brown Green/Yellow Blue Mains Cord with Mains Plug Mains Cord Ferrite B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
The digital board is always delivered with the latest LLS/LLC software versions! 34560734 Analog Board AB (38108641) 34560866 HDF Online Board HOB (38108647) 3451665A MSBK Board (3810671X) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 403
10 Fuse T3.15 A (TR5) LH1 LH2 N PE PE K3/K4 34514775 2 Relay (12 V) for 24 V C114/C115 34519912 2 Capacitor 1F H104 34560580 1 Buzzer 34560416 10 Plug for buzzer B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
LH1 LH2 N PE PE 7.7.3.1 Compatibility Table SMPS-MC SMPS-MC Type Plate Version Number SMPS-MC SW Version Dialog+ SW 9.xx (TN/Item-No) (Ver/SW) XXXXXX.5 Rev. 1.15 34560220 (version 5) ex works B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 405
1 Metal disc ∅ 10 mm for cover 9 34560033 9 34519823 Hinge pin with interior thread hinge pin 9 34516670 1 Seal for blood pump SAKV-SG SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
17 34560831 see separate figure (paragraph 7.8.5) 1 Valve block level regulation (double pump) 18 34560823 1 Diaphragm pump with assembly elbow 3 mm 4 mm 4 mm SAKA SAKV-SG B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 Reed sensor with contacts and o-ring 3.68 x 1.78 - 34516670 1 Seal for blood pump ^^^^^^^^^^^^^^^^^^^^^^^^ 100 g tube (to lubricate the shaft of the blood pump) - 7703708 1 Grease SKD 4002 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
SAD with board; do not tamper, exchange only complete red detector If cover for SAD is present: a ball head allen key 2.5/3 is required for bottom screw behind rotary magnet. SAKA SAKV-SG B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
3m Silicone tubing 8 x 3.2 / red (3 m), external high temperature tubing with two single ear clamps 17.5 - 39224538 1 Single ear clamp 17.5 for external high temperature tubing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 Male connector (red) with o-rings 8 34570004 5 O-ring 9.3 x 2.4 9 34565280 3m Silicone tubing 3.2 x 1.6 (3 m) 10 3451728A 1 PE tube 4 x 1 with silicone tubing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Comments Central Concentrate Supply (Option) 3 34565248 10m Silicone tubing 5 x 1.5 / blue (10 m) 5 34565230 10m Silicone tubing 5 x 1.5 / red (10 m) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
13 34511792 1 Tubing coupling 14 34517138 1 Reed sensor with contacts and o-ring 3.68 x 1.78 15,14,13 12,11,8 15 34771255 20 O-ring 14 x 1.5 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 418
10 O-ring 5.0 x 1.6 for eccecntric shaft 6 34560858 1 Eccecntric roller 7 34560840 1 Eccecntric shaft 8 3451659D 1 Servomotor with board for VD, servomotor and board must be exchanged together B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 419
17 34570730 1 Housing complete for temperature sensor 18 34773762 10 O-ring 9.25 x 1.78 19 34511792 1 Tubing coupling 20 34771255 20 O-ring 14 x 1.5 20,19 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 422
* Balance (e.g. Pocket Balance) For example manufacturer: for details contact manufacturer Kern & Sohn GmbH directly Internet: www:kern-sohn.com * not in the product range of B. Braun Avitum AG B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 Male to male connector couplings for the measurement of the patient leakage current - 770284A 1 DCI cable, external required for motor assembly - 7703848 1 Assembly jig (motor for piston pump) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Comments 7.12 Service Kits - 3451897A 1 Repair kit balance chamber for the complete replacement of both membranes, incl. all o-rings and assembly parts (self-locking nuts, screws, springs etc.) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 CD Remote Bedside Link List ReBeLL, for network version 1.5.13 - 7703562 for WAN-BSL only 1 CD WAN-BSL, firmware version 1.0.4.8 - 7703864 1 Field service utility FSU, version 2.00 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 426
Software 7.58 (Linux) Software 7.59 (Linux) *1.9 Software-Update 7.59 (Linux) #1.01/ USB Stick with TLC/LLS SW only 1.04 if SW 7.52 is present Software ≥ 8.20 (Linux) Software 9.02 (Linux) 1.04 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 PVC tubing 10 x 3 / black, (3 m), with a single ear clamp 19.5 and a tubing clamp 12-20 mm - 39234126 1 Tubing clamp 12-20 mm - 34565345 1m Silicone tubing 12 x 2.5 (1 m) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 428
- 7703708 1 Grease SKD 4002 100 g tube (to lubricate the shaft of the blood pump) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 Substitution connector 5 3477106A 5 O-ring 6.4 x 1.3 6 34517138 1 Reed sensor with contacts and o-ring 3.68 x 1.78 A or A+B (A=Quadring; B=Spacer) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
1 34560955 1 Filter (ABPM module M2009) 2 3451487C 1 ABPM module M2009 firmware version of ABPM module on sticker: Version Y (FW):X.X with filter Version YX.X Version YX.X B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Description Comments 7.19.1 Compatibility Table ABPM Module M2009 Position of DIL Switch for DIL Switch: Default LX800 (ex Works) LX800 Motherboard DIL Switch DIL Switch Version YX.X LX800 (default) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 440
Sulfric acid in the eyes may cause loss of eyesight and acid on the skin will cause burns. Pay attention to the included documentation of the battery supplier! see separate figure - Battery sub-rack B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 441
F301 H302 F302 H304 F304 H601 F601 brown blue brown H602 F602 H402 F402 F403 H403 blue green/yellow green/yellow brown blue GNDAKKU XP3.2 brown brown XP3.1 +24VAKKU P9 P7 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 443
7 . S p a r e P a r t s L i s t 1 / 2 0 1 0 7 - 69 SW 9.xx Art. No. Qty. Description 7.22 Card Reader (Option) 1 34560548 1 Card reader board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 444
7 . S p a r e P a r t s L i s t 1 / 2 0 1 0 7 - 70 SW 9.xx Art. No. Qty. Description 7.23 Adimea Option 1 3456063A 1 Kt/V-UV board with housing and tubing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd...
Page 445
8 - 1 SW 9.xx ESD/EMC Information 8.1.1 Electrostatic Discharge ESD 8.1.2 Electromagnetic Compatibility EMC Technical Information TI Assembly Instructions AI Field Service Information FSI Instruction Leaflets IL B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd...
Page 446
In general the mats are made of sturdy two layer material and have reinforced Mat Material edges and corners. The work surface is static dissipative (> 100 MΩ), the other side is conductive (> 100 kΩ). B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd...
Page 447
Unprotected electronic parts (e.g. wrapped in normal plastic film or parts padded with foam material) are assumed to be damaged by electrostatic discharge and will be scrapped. B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd...
Page 448
N, connected to a one or three phase public power supply with a constant current load of the power supply ≥ 100 A per phase, then it is also valid for the Dialog machine with the switch mode power supply microcontroller SMPS-MC. B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd...
Page 449
The power frequency magnetic field should be measured in the intended installation location to magnetic field acc. to IEC 61000- assure that it is sufficiently low. NOTE U is the a.c. mains voltage prior to application of the test level. B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd...
Page 450
Dialog machine. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd...
Page 451
Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 - 7 SW 9.xx Technical Information TI B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd...
Page 452
Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 - 8 SW 9.xx Assembly Instructions AI B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd...
Page 453
Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 - 9 SW 9.xx Field Service Information FSI B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd...
Page 454
Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 - 10 SW 9.xx Instruction Leaflets IL B. Braun Avitum AG BA-TE-DE08C M.KAY - Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd...
Page 455
2011-03-01 4-1-1_1/2010 Revision Chapter 4 TSM Service Program 2011-03-01 4-2-1_1/2010 4-3-1_1/2010 5-1_1/2010 Revision Chapter 5 Technical Safety Inpsection and Preventive Maintenance 2011-03-01 9-1_1/2010 Revision Chapter 9 Edition/Updates 2011-03-01 B. Braun Avitum AG BA-TE-DE08W M.KAY - Dialog+ SW9xx_SM_Chapter 9-1_1-2010.doc/pdf <110301> yymmdd...
Need help?
Do you have a question about the Dialog+ and is the answer not in the manual?
Questions and answers
ola tenho varias placas da bbraum que gera a imagem pro monitor modelo: BBM-LX800-A1R ele simplesmente ao ser ligada não bipa, e consequetemente nao funciona nada! so ascende um led na placa e mais nada. gostaria de saber se o defeito pode ser bios, memoria chipset, ou o processador que no caso e o AMD GEODE. desde de ja agradeço
The cause of the B. Braun Dialog+ board not beeping or functioning when connected to a BBM-LX800-A1R monitor, with only one LED lighting up, could potentially be related to the BIOS, memory, chipset, or AMD GEODE processor. However, based on the context provided, the most likely causes to investigate include:
1. BIOS Issue: Ensure the BIOS is functioning properly and has the correct version. Refer to the procedure for checking the BIOS version as outlined in the context (e.g., pressing the TAB key during startup to display the BIOS version). A corrupted or incompatible BIOS could prevent the board from functioning.
2. Memory Issue: Faulty or improperly seated memory modules can cause similar symptoms. Verify that the memory is correctly installed and functional.
3. Chipset Issue: A malfunction or damage in the motherboard chipset could also result in the board failing to operate.
4. AMD GEODE Processor Issue: If the processor is damaged or not seated correctly, it could prevent the board from functioning.
Careful diagnostics are needed to isolate the specific cause.
This answer is automatically generated