Pressure Reduction - B. Braun Dialog+ Service Manual

Sw 9.xx
Table of Contents

Advertisement

+
Dialog
SW 9.xx

3.3.8.13 Pressure Reduction

FPE
RVDA
Fig. : Flow Diagram Pressure Reduction
LLC Test
Code
1280x
13.
VEBK test
FPE = 500 ml/min
FPA = 500 ml/min
Valve Status:
VEBK1/2, VDEBK1/2 ------------------- opened
VABK1/2, VDABK1/2------------------- opened
VBP, VDA, VDE ------------------------ opened
Timeout 60 s
BA-TE-DE08C M.KAY
Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
3 . R e p a i r I n s t r u c t i o n s
VDEBK
VEBK
A pressure level is set at PDA between –200 mmHg and +200 mmHg by
opening all valves and driving the pumps accordingly.
LLC Test Steps
FPA
VDABK
VABK
LLS Test Steps
1 / 2 0 1 0
VDE
VBP
VDA
PDA
open
UFP
offen
closed
geschlossen
B. Braun Avitum AG
3 - 28
LLS Test
Code
12000

Hide quick links:

Advertisement

Chapters

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Dialog+ and is the answer not in the manual?

Questions and answers

Eduardo Felipe de Paula
January 22, 2025

ola tenho varias placas da bbraum que gera a imagem pro monitor modelo: BBM-LX800-A1R ele simplesmente ao ser ligada não bipa, e consequetemente nao funciona nada! so ascende um led na placa e mais nada. gostaria de saber se o defeito pode ser bios, memoria chipset, ou o processador que no caso e o AMD GEODE. desde de ja agradeço

1 comments:
Mr. Anderson
February 10, 2025

The cause of the B. Braun Dialog+ board not beeping or functioning when connected to a BBM-LX800-A1R monitor, with only one LED lighting up, could potentially be related to the BIOS, memory, chipset, or AMD GEODE processor. However, based on the context provided, the most likely causes to investigate include:

1. BIOS Issue: Ensure the BIOS is functioning properly and has the correct version. Refer to the procedure for checking the BIOS version as outlined in the context (e.g., pressing the TAB key during startup to display the BIOS version). A corrupted or incompatible BIOS could prevent the board from functioning.

2. Memory Issue: Faulty or improperly seated memory modules can cause similar symptoms. Verify that the memory is correctly installed and functional.

3. Chipset Issue: A malfunction or damage in the motherboard chipset could also result in the board failing to operate.

4. AMD GEODE Processor Issue: If the processor is damaged or not seated correctly, it could prevent the board from functioning.

Careful diagnostics are needed to isolate the specific cause.

This answer is automatically generated

Table of Contents