Dialog
SW 9.xx
Disinfection after each Therapy
Automatic Preparation Start
Thermal Disinfection
Chemical Disinfection
Short Chemical Disinfection
Central Thermal Disinfection
Central Chemical Disinfection
Auto Switch-On with Rinsing
- No Weekly Schedule
BA-TE-DE08C M.KAY
Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
4 . T S M S e r v i c e P r o g r a m
If the YES key is activated, a disinfection must be performed after every
therapy.
YES
After disinfection and rinsing the Preparation and the self tests are started.
after Disinfection
NO
After disinfection and rinsing the disinfection method is terminated.
You can enable the thermal disinfection with the YES key in the disinfection
program or disable with the NO key. If you select the NO key the thermal
disinfection icon is not displayed in the disinfection program.
You can enable the chemical disinfection with the YES key in the disinfection
program or disable with the NO key. If you select the NO key the chemical
disinfection icon is not displayed in the disinfection program.
You can enable the short chemical disinfection with the YES key in the
disinfection program or disable with the NO key. If you select the NO key the
short chemical disinfection icon is not displayed in the disinfection program.
The inlet feed-line and the machine are disinfected with hot water, if the
central facility loop-line is also disinfected with hot water (> 85
The inlet feed-lines are disinfected with chemical disinfectants, if the central
facility loop-line is also disinfected with chemical disinfectants.
Rinsing of disinfectant residues: e.g. a disinfection was performed in the
Rinsing
evening and the residues are rinsed in the morning.
1 / 2 0 1 0
The parameters for the switch-on without a
weekly schedule can be set in the menu Auto
Switch-On with Rinsing - No Weekly Schedule.
The keys for Auto Switch-On with Rinsing - No
Weekly Schedule and Auto Switch-On by Weekly
Schedule are interlocked.
The machine is rinsed after the preselected
switch-on time. The key is only enabled if the
function is selected.
Switch-On Time
The switch-on time can be set (without a weekly
schedule).
4 - 40
o
C).