Uf Parameters - B. Braun Dialog+ Service Manual

Sw 9.xx
Table of Contents

Advertisement

+
Dialog
SW 9.xx
4.7.1.8

UF Parameters

UF Rate Compensation?
Max. Value of Upper Limit UF Rate
Default Value of Upper Limit UF Rate
Gross UF Rate vs. Blood Flow -
BA-TE-DE08C M.KAY
Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
4 . T S M S e r v i c e P r o g r a m
You can set the minimal UF rate.
Minimal UF-Rate
If YES is Activated
If the Minimal UF Rate is active the UF rate is recalculated to reach the
required UF volume.
You can set the maximum value of the upper limit for the UF rate.
You can set the default value of the upper limit for the UF rate.
You can set the UF volume.
UF Volume
You can set the therapy time.
Therapy Time
You can set and activate an alarm limit, i.e. an alarm (red) is activated if the
limit is reached. The alarm limit can only be changed if the warning limit was
Alarm Limit
changed in advance. The alarm can be acknowledged with the
Setting Range
The alarm limit can not be set lower than the warning limit (even if the
warning limit is deactivated).
1 / 2 0 1 0
The dialysate fluid parameters can be set in the UF
Parameters menu.
1. Activate the System Configuration menu with
the System Configuration key in menu
Treatment Support. The following menu is
opened and displayed (scroll with scroll bar).
2. Activate the UF Parameters menu with the UF
Parameters icon. The following menu is
opened.
The following parameters can be set:
• Minimal UF Rate
• UF Rate Compensation?
• Max. Value of Upper Limit UF Rate
• Default Value of Upper Limit UF Rate
• UF Volume
• Therapy Time
• Gross UF Rate vs. Blood Flow - Alarm Limit
• Gross UF Rate vs. Blood Flow - Warning Limit
• UF Profile Editor
4 - 35
key.
B. Braun Avitum AG