Dialog
SW 9.xx
4.1
Start TSM Service Program
Basic Board
BA-TE-DE08C M.KAY
Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
4 . T S M S e r v i c e P r o g r a m
TSM Service Program
HDF
Power Board Valves
Digital Board
Analog Board
Only activate the TSM service program for
service activities. It is prohibited to connect a
patient to the Dialog+ and to run a therapy if
the TSM service program is activated in the
Dialog+. If the TSM service program is
activated the complete alarm system is
disabled.
Therapy Mode
After completion of all procedures switch back
to the therapy mode: digital board, service
switch S1, position 0.
The TSM Main Menu (TSM = Technical Support
and Maintenance) is used for servicing the
machine. The TSM service program (TSM Main
Menu) is started as follows:
1. Switch off machine.
2. Open rear door.
3. On digital board DB: turn service switch S1
from position 0 to position 2.
Service Switch S1:
Position 0 = Therapy
Position 2 = TSM Service Program
Position 3 = Software Installation/Update
4. Switch on machine.
The machine starts in TSM mode.
Online
Board
JP1
P1 1
1 / 2 0 1 0
Power Board
M otors
JP2
S1
B. Braun Avitum AG
4 - 5
Need help?